Evidence map›Paper›PMID 41858738›Full record

ArticleJournal of vascular surgery cases and innovative techniques2026

Irretrievable wire wrap and catheter entrapment: Novel failure mode of endobypass device.

Shina Patel, Jocelynne T Dorotan, Kenny Oh

Abstract readCase Reports
In one paragraph

Article in Journal of vascular surgery cases and innovative techniques, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Shina PatelDepartment of Vascular Surgery, Lewis Katz School of Medicine, Philadelphia, PA.
Jocelynne T DorotanDepartment of Vascular Surgery, Lewis Katz School of Medicine, Philadelphia, PA.
Kenny OhDepartment of Vascular Surgery, Lewis Katz School of Medicine, Philadelphia, PA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

A 75-year-old man with a history of severe ischemic cardiomyopathy, coronary artery disease status post coronary artery bypass graft, stroke, atrial fibrillation, chronic kidney disease, and recent right metatarsophalangeal amputation was admitted for an ischemic nonhealing wound. Angiography demonstrated right superior femoral artery occlusion with above-knee reconstitution. Percutaneous transmural arterial bypass using the Detour device (Endologix, LLC) was complicated by severe wire wrap and irretrievable catheter entrapment, necessitating open extraction and procedure abortion. Despite technical failure, the wound had healed at follow-up. This case highlights device-specific challenges and divergence between technical and clinical success.

Indexed as

Detour endobypass systemDevice complicationEndovascular interventionPeripheral arterial disease

Identifiers

PMID41858738
PMCPMC12996263

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.