ArticleOphthalmology and therapy2026
Validation of the Eyerobo FC Portable Fundus Camera for Diabetic Retinopathy Screening Using Public Datasets and Deep Learning.
Article in Ophthalmology and therapy, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Machine Learning-Based Prediction of Cycloplegic Refraction Using the Eyerobo Vision Screener: Design-A Cross-Sectional Comparative Device Study.Ophthalmology and therapy · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors.
Funding
Abstract
introductionTo validate the diagnostic performance of the Eyerobo FC, a new portable non-mydriatic fundus camera for diabetic retinopathy (DR) screening, against an established desktop fundus camera benchmark using a transfer-learning approach in which artificial intelligence (AI)-based detection algorithms trained on desktop images were applied to Eyerobo FC images.
methodsThis prospective validation study employed a three-tier experimental design. Tier 1 involved training a deep learning model (EfficientNet-B4) on standard desktop camera images from EyePACS and APTOS 2019 datasets. Tier 2 established the reference standard by evaluating the trained model on the Messidor-2 dataset (N = 1748 eyes) captured with a Topcon TRC NW6 desktop camera (sensitivity 92.7%, 95% CI 91.2-94.2%; AUC 0.952, 95% CI 0.943-0.961). Tier 3 validated the same AI model (without retraining) on images from the Eyerobo FC in a prospective cohort (N = 104 eyes: 52 referable DR, 52 non-referable). The primary outcome was noninferiority of sensitivity and specificity (margin
resultsIn a balanced cohort of 104 eyes (52 referable DR, 52 non-referable), the Eyerobo FC achieved sensitivity of 92.3% (95% CI 84.4-98.2%) and specificity of 94.2% (95% CI 87.0-100%), demonstrating noninferior performance compared with the desktop benchmark (sensitivity 92.7%, specificity 94.3%). The sensitivity difference of
conclusionsThe Eyerobo fundus camera demonstrates noninferior diagnostic performance (sensitivity 92.3%, specificity 94.2%, AUC 0.977) compared with desktop systems when evaluated with AI algorithms trained exclusively on desktop images. These findings support deploying portable AI-assisted screening in resource-constrained and point-of-care settings, with successful cross-domain transfer learning enabling algorithmic generalizability across imaging platforms.
Indexed as