Evidence map›Paper›PMID 41810137›Full record

ArticleNorth American Spine Society journal2026

Ten-year analysis of 17,639 thoracolumbar instrumentation failures using the manufacturer and user facility device experience (MAUDE) database.

Gil Kimchi, Darcy Johnston, John Paul G Kolcun, Jill Serbousek, Richard G Fessler

Abstract read
In one paragraph

Article in North American Spine Society journal, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Gil KimchiDepartment of Neurosurgery, Rush University Medical Center, 1725 W. Harrison St., Chicago, IL 60612, United States.
Darcy JohnstonIndependent Data Scientist, In Queue Innovations, PO Box 420, Winona Lake, IN 46590, USA.
John Paul G KolcunDepartment of Neurosurgery, Rush University Medical Center, 1725 W. Harrison St., Chicago, IL 60612, United States.
Jill SerbousekIn Queue Innovations, PO Box 420, Winona Lake, IN 46590, USA.
Richard G FesslerDepartment of Neurosurgery, Rush University Medical Center, 1725 W. Harrison St., Chicago, IL 60612, United States.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: This study utilized the FDA-maintained Manufacturer and User Facility Device Experience (MAUDE) database, an extensive resource capturing medical device reports (MDRs) of adverse events, to perform a cross-sectional analysis of failure mechanisms reported by manufacturers. Methods: All thoracolumbar pedicle screw system failures reported in the MAUDE database between January 2014 and December 2023 were identified using FDA product codes. Failures were classified as intraoperative or postoperative and grouped into 9 and 5 clusters, respectively. Following de-identification, the 8 most frequently reported manufacturers were analyzed individually. Reports specifying failed components were categorized as screw, set screw, rod, instrument, system, other, or unknown. The "system failure" category lacked specificity regarding the failed component. Descriptive statistics summarized failure mechanisms. Results: A total of 17,639 MDRs related to thoracolumbar fixation failures were reported over the past decade, with 15,234 postoperative and 2,405 intraoperative events. The most frequent intraoperative issues were positioning problems (27.9%), device-device compatibility (26.4%), and difficulty in separation/removal (11.8%), while the most common postoperative failure mechanisms were breakage (45.2%), migration/detachment (42%), and deformation (10.1%). The distribution of failed components varied by the associated failure mechanism. Among 4,676 reported breakages, screws (20.8%), set screws (15.5%), and rods (11.8%) were most frequently involved. In the 4,871 migration/detachment events, set screws accounted for 31.6%, screws for 21%, and rods for 2.3%. Of 1,325 deformation events, the majority were linked to set screws (76.5%), with fewer attributed to system-level components (11.2%) and screws (6%). Conclusions: Failure profiles for pedicle screw systems vary significantly between manufacturers. The MAUDE database is a useful resource for the evaluation of reporting patterns associated with pedicle screw systems. However, limitations include reporting biases and the absence of total sales data, which prevents the calculation of true failure rates.

Indexed as

FailureFDALumbar fixationMalfunctionMAUDEPedicle screwsThoracolumbar fusion

Identifiers

PMID41810137
PMCPMC12969441

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.