Multileaf Faraday cup for beam energy verification in radiation therapy with ultra-high dose-rate electron beams and ion beams.
Article in Medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
Abstract
backgroundIn radiotherapy and particle therapy, the stability of the beam energy must be checked routinely during quality assurance. This is of particular importance when varying the dose rate via the beam intensity of the accelerator in order to study the FLASH effect when comparing irradiations with conventional and ultra-high dose rates. PURPOSE: The standard method of energy verification based on the measurement of percent depth-dose curves in a water phantom is time-consuming and in the case of ultra-high dose-rate beams, may also present a non-negligible radiation protection problem.
methodsA compact, portable multileaf Faraday cup (MLFC) with 128 channels was developed for energy determination based on the measurement of depth-charge curves. Its design is optimized for clinical electron beams in a range between 3 and 25 MeV. The read-out unit of the MLFC displays the beam energy in real time. The device was tested in strongly pulsed electron beams such as those present in irradiations with ultra-high doses per pulse as well as in conventional clinical electron beams. The same detector was used for proof of concept in a single measurement campaign in pulsed beams of therapeutical protons (conventional dose rates) from a synchrotron source using a suitable range shifter. The MLFC was calibrated with monoenergetic electron beams as well as against depth-dose curves. Simulations were carried out for comparison.
resultsThe MLFC works well under ultra-high pulse dose rate conditions as well as in conventional electron beams generated by a medical accelerator. Changes in the beam energy of below 20 keV (0.1% of 20 MeV) can be clearly identified by means of the MLFC. For proton beams, well-defined peaks in the depth-charge curves are observed for each single synchrotron spill with a total charge of only about 1 nC. The energy values resulting from MC simulations of the measured MLFC data agree with the actual proton energies within 2
conclusionsThe MLFC can be used for quick validation of the energy stability when carrying out experiments comparing electron beams of conventional and ultra-high dose rates. The same device can be used in the pulsed ion beams from a synchrotron.