To shift or not to shift: identifying and correcting patient motion after couch rotations in non-coplanar intracranial radiosurgery with stereoscopic X-ray imaging.

Vanessa Da Silva Mendes, Sylvia Garny, Lili Huang, Stephan Schönecker, Christian Trapp, Frederik Fuchs, Christopher Kurz, Claus Belka, Guillaume Landry, Michael Reiner and 1 more

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

11 authors.

Vanessa Da Silva MendesDepartment of Radiation Oncology, LMU University Hospital, Munich, Bavaria, Germany.ORCID https://orcid.org/0000-0002-6706-8590
Sylvia GarnyDepartment of Radiation Oncology, LMU University Hospital, Munich, Bavaria, Germany.
Lili HuangDepartment of Radiation Oncology, LMU University Hospital, Munich, Bavaria, Germany.
Stephan SchöneckerDepartment of Radiation Oncology, LMU University Hospital, Munich, Bavaria, Germany.
Christian TrappDepartment of Radiation Oncology, LMU University Hospital, Munich, Bavaria, Germany.
Frederik FuchsDepartment of Radiation Oncology, LMU University Hospital, Munich, Bavaria, Germany.
Christopher KurzDepartment of Radiation Oncology, LMU University Hospital, Munich, Bavaria, Germany.
Claus BelkaDepartment of Radiation Oncology, LMU University Hospital, Munich, Bavaria, Germany.
Guillaume LandryDepartment of Radiation Oncology, LMU University Hospital, Munich, Bavaria, Germany.ORCID https://orcid.org/0000-0003-1707-4068
Michael ReinerDepartment of Radiation Oncology, LMU University Hospital, Munich, Bavaria, Germany.
Stefanie CorradiniDepartment of Radiation Oncology, LMU University Hospital, Munich, Bavaria, Germany.

Funding

Brainlab SE
6 · The paper itself

Abstract

backgroundFrameless linear accelerator (linac)-based image-guided stereotactic radiosurgery (SRS) or fractionated stereotactic radiotherapy (FSRT) are a widely used treatment option for intracranial lesions. Given the high radiation doses involved, it is crucial to maintain precise patient positioning throughout treatment. This requires that geometric inaccuracies arising from patient motion or setup errors are identified and corrected. With frameless immobilization, image-guidance has a greater impact, especially in non-coplanar settings that can lead to patient motion and discrepancies between couch and radiation isocenters. PURPOSE: Both patient and phantom studies were conducted to assess and quantify the magnitude of geometric uncertainties after couch rotations, aiming at evaluating the clinical need for their correction to warrant a precise treatment delivery.

methodsIntrafraction X-ray data, performed by ExacTrac Dynamic (ETD) to monitor and correct patients' position throughout treatment delivery, were collected from 50 patients treated for brain metastases in stereotactic non-coplanar schemes and immobilized by stereotactic double-layered thermoplastic mask systems: 26 patients treated in 40 single-fraction SRS (168 stereoscopic X-ray images); 24 treated with FSRT in 128 fractions (278 stereoscopic X-ray images). Additionally, a head phantom was utilized and 350 measurements under two different couch loads were carried out to distinguish true patient motion from deviations caused by couch rotations or system-related effects. For both studies, ETD stereoscopic X-rays were acquired after each couch rotation and the first measured positioning deviation was calculated by comparing X-ray images to the treatment plan's digitally reconstructed radiographs.

resultsClinically relevant deviations were observed, exceeding clinical tolerance (≥ 0.5 mm/0.5°) mostly in the lateral and yaw directions and requiring repositioning in nearly half of the couch rotations. These deviations measuring up to 2 mm, revealed to be emerging mainly from patient motion rather than linac setup, as the phantom study showed maximum deviations of up to 0.6 mm and 0.4° when simulating a patient treatment and an interquartile range that did not exceed 0.2 mm and 0.2°.

conclusionsThese findings demonstrate the importance of a continuous intrafraction motion monitoring and repositioning in cranial stereotactic treatments, especially in non-coplanar settings.

Indexed as

Brain NeoplasmsPatient PositioningPhantoms, ImagingRadiosurgeryRadiotherapy, Image-GuidedRadiotherapy Planning, Computer-AssistedRadiotherapy Setup ErrorsTomography, X-Ray ComputedHumansImmobilizationMovementParticle AcceleratorsRadiotherapy DosageRadiotherapy, Intensity-ModulatedRotationIGRTintracranialnon‐coplanar beamspatient (re‐)positioningSRSstereoscopic X‐ray imagingX‐ray intrafraction monitoring

Identifiers

PMID41795648
PMCPMC12967575

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.