Evidence map›Paper›PMID 41772588›Full record

ArticleMalaria journal2026

Performance evaluation of an automated method for accurate identification of glucose-6-phosphate dehydrogenase deficiency on the BS-2800M analyzer.

Qinlan Liu, Anping Xu, Yueying Huang, Feng Wang

Abstract readEvaluation Study
In one paragraph

Article in Malaria journal, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Qinlan LiuClinical Laboratory, Shenzhen Baoan Women's and Children's Hospital, Shenzhen, 518100, China.
Anping XuDepartment of Laboratory Medicine, Peking University Shenzhen Hospital, Shenzhen, 518036, China.
Yueying HuangDepartment of Laboratory Medicine, Peking University Shenzhen Hospital, Shenzhen, 518036, China.
Feng WangClinical Laboratory, Shenzhen Baoan Women's and Children's Hospital, Shenzhen, 518100, China. wangfengnew@qq.com.

Funding

Zhongnanshan Medical Foundation of Guangdong Province ZNSXS-20240021
6 · The paper itself

Abstract

The measurement of glucose-6-phosphate dehydrogenase (G6PD) activity was currently used laboratory method for the diagnosis of G6PD deficiency (G6PDd). This study evaluated the analytical and diagnostic performance of a fully-automated method for the one-time quantitative measurement of G6PD activity normalized per unit of hemoglobin (G6PDH) on the Mindray BS-2800M analyzer, using 1561 whole blood samples. For analytical performance, the G6PDH (U/g Hb) showed more stable repeatability performance with coefficient of variation (CV) of 4.48% ± 3.86%, compared with the relatively large CVs in G6PD and Hb detection alone. G6PDH exhibited linearity from 0.37 to 20.86 U/g Hb, which could almost cover the known very low level and very high level, with a carryover rate of less than 1%. The minimum applicable level of Hb for blood sample was at 200 g/L, in which the G6PDH automated method results were not be interfered significantly compared with manual method. Sample stability studies indicated that G6PDH results remained stable for up to 4 h at room temperature and 48 h at 4 °C, but were significantly reduced after storage at - 20 °C (P < 0.05). To evaluate the diagnostic performance of G6PDH, the reference intervals in 806 non-neonates were established as < 2.31 U/g Hb for deficient males/females, 2.31-6.17 U/g Hb for intermediate females. For neonates, they were < 4.14 and 4.14-11.03 U/g Hb respectively based on 122 samples. Comparison of genetic results from 185 samples revealed that G6PDH achieved accuracy of 91.35%, surpassing conventional G6PD single enzymatic method (77.30%) and G6PD/6PGD (82.70%). The kappa coefficient and chi-square analysis indicated that G6PDH phenotypic classifications and genetic results had significant correlation. Furthermore, in 64 of heterozygous females, G6PDH had higher positively detection rate (87.50%) than G6PD single enzymatic method (39.06%) and G6PD/6PGD (51.56%). In summary, the automated G6PDH method offered a valuable approach for clinical laboratory, as it not only had favorable analytical performance for improving testing efficiency, but also provided reliable diagnostic performance for G6PDd, especially holding significant value in precise screening female heterozygotes, early prevention of G6PDd and guiding safe clinical medication.

Indexed as

Automation, LaboratoryGlucosephosphate DehydrogenaseGlucosephosphate Dehydrogenase DeficiencyFemaleHumansReproducibility of ResultsSensitivity and SpecificityGlucosephosphate DehydrogenaseAnalytical performanceAutomated lysisDiagnosis efficacyGlucose-6-phosphate dehydrogenase deficiencyMethod comparison

Identifiers

PMID41772588
PMCPMC13072469

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.