In Vitro Hydrodynamic Evaluation of Pulmonary Expanded Polytetrafluoroethylene Valved Conduits.
Article in Interdisciplinary cardiovascular and thoracic surgery, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- RE: In Vitro Hydrodynamic Evaluation of Pulmonary Expanded Polytetrafluoroethylene Valved Conduits.Interdisciplinary cardiovascular and thoracic surgery · 2026Article
- Reply to Hinpetch Daungsupawong et al.Interdisciplinary cardiovascular and thoracic surgery · 2026Article
Corrections and comments
- Commented on by
Authors and funding
7 authors.
Funding
Abstract
objectivesVarious designs are proposed for pulmonary expanded polytetrafluoroethylene (ePTFE) conduits and have been applied in clinical practice. However, experimental data to support them are limited. We conducted an in vitro experiment using a circulatory simulator to evaluate their haemodynamic performance and hydrodynamic characteristics.
methodsThree root models with a 24-mm basal ring (A, straight; B, with small sinuses; C, with large sinuses) were 3D-printed. Cusps were uniformly cut out from a 0.1-mm-thick ePTFE membrane and sewn to the inter wall of the models. Model A had a single suture on the free margins near the commissures. Each model was tested with a pump size of 70 mL, 70 beats/min, and arterial pressure of 30/10 mmHg. The valve behaviour was recorded by a high-speed camera, and particle image velocimetry (PIV) was performed in the region behind the model housing section.
resultsPeak transvalvular pressure gradients were 4.0, 4.8, and 4.3 mmHg (P = .95), and geometric orifice areas were 2.34, 2.38, and 2.46 cm2 (P = .96) in models A, B, and C, respectively. Particle image velocimetry revealed peak instantaneous velocity was 1.69, 1.69, and 1.65 m/s (P = .74) and peak Reynolds shear stress in the midsystolic phase was 56.8, 49.5, and 25.5 Pa (P = .05) in models A, B, and C, respectively. Model C tended to have a lower distribution of turbulent flow than the other models.
conclusionsAll models exhibited sufficient opening and acceptable Reynolds shear stress values. The sinus contributed to the suppression of turbulent flow, which may lead to an improvement of conduit durability, but its effect was dependent on the sinus size.