ArticleWideochirurgia i inne techniki maloinwazyjne = Videosurgery and other miniinvasive techniques2025
Mechanical stability of new‑generation meshes for M3 inguinal hernia repair: experimental pressure chamber testing of SWING‑Mesh and 3DMax MID Anatomical Mesh.
Article in Wideochirurgia i inne techniki maloinwazyjne = Videosurgery and other miniinvasive techniques, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT07289802 (Safety and Efficacy of the 3DMax MID Anatomical Mesh), which is not on this map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Safety and Efficacy of the 3DMax MID Anatomical Mesh (BD, USA) in Laparoendoscopic Inguinal Hernia Repair: A Multicenter Prospective Observational Cohort Study
Who cites it
1 citing paper in PubMed.
- Early safety and effectiveness of new 3‑dimensional lightweight SWING‑Mesh in minimally‑invasive inguinal hernia repair: a multicenter prospective observational study with 3‑month follow‑up.Wideochirurgia i inne techniki maloinwazyjne = Videosurgery and other miniinvasive techniques · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionThe necessity of mesh fixation in laparoendoscopic repair of large medial (classified by the European Hernia Society as M3) inguinal hernias (IHs) remains debated. Recent data, including the MEFISTO randomized controlled trial, suggest that rigid, anatomically contoured meshes may provide sufficient stability of hernia repair without mesh fixation, potentially reducing fixation-related complications. However, biomechanical performance of newly introduced anatomical meshes has not been thoroughly evaluated.
aimThis study aimed to assess the mechanical stability of 2 newly introduced anatomically shaped meshes-SWING-Mesh and 3DMax MID Anatomical Mesh-in a validated pressure chamber model simulating M3 IH defects. MATERIALS AND
methodsA rigid 3-dimensionally (3D) printed groin model with a 4-cm medial defect was mounted in a sealed pressure chamber capable of generating intra-abdominal pressures of up to 70 kPa. Two meshes were tested: SWING-Mesh (lightweight polypropylene; 55 g/m²; 16 cm × 12 cm) and 3DMax MID Anatomical Mesh (medium-weight polypropylene; 78 g/m²; 17 cm × 12 cm). Both were positioned with an at least 3-cm overlap beyond the defect margins without fixation. Initial testing was performed at 36 kPa. If displacement occurred, lower pressures were applied; if stability was maintained, higher overload pressures were used. Each experiment was repeated 3 times, with outcomes documented on photo and video material. The primary end point was mesh displacement into the defect.
resultsSWING-Mesh consistently displaced into the defect at 36 kPa in all trials. Supplementary tests showed migration as early as at 10 kPa. In contrast, 3DMax MID Anatomical Mesh remained stable at 36 kPa in all repetitions. Overload testing confirmed its resistance at 53, 60, and 70 kPa-the maximum achievable pressure in the chamber-without measurable displacement.
conclusionsSWING-Mesh failed to provide stability even under moderate intra-abdominal pressures, which questions its suitability for nonfixed repair of M3 hernias. 3DMax MID Anatomical Mesh demonstrated complete stability under both physiologic and supraphysiologic conditions, confirming the mechanical advantage of medium-weight 3D meshes. Implant design, weight, and stiffness appear more decisive for stability than fixation, supporting the selective use of nonfixation techniques when rigid meshes are applied.