ArticleJournal of cardiovascular electrophysiology2026
Real-Time 3D Mapping Corrects Fluoroscopic Misperception During Pulmonary Vein Isolation by Pulsed Field Ablation.
Article in Journal of cardiovascular electrophysiology, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundPulsed field ablation (PFA) for pulmonary vein isolation (PVI) is mainly guided by fluoroscopy. A novel electroanatomical module enables real-time 3D mapping (3D-EAM) integrated with the PFA catheter.
objectivesTo assess the contribution of 3D-EAM to pulmonary vein (PV) targeting accuracy and lesion coverage during PFA.
methodsAmong 66 consecutive prospective cases, two sequential phases were conducted: a 3D-blinded mapping phase (n = 21), in which anatomical maps were created but withheld from operators, and a real-time 3D-EAM guidance phase (n = 45), where live mapping was used during ablation. A retrospective cohort of fluoroscopy-only PFA procedures (n = 77) served as a control. Lesion coverage and procedural metrics were compared across groups, and post-procedural operator assessments were collected in the 3D-blinded phase.
resultsIn the 3D-blinded phase, 5 of 21 procedures (23.8%) failed to isolate at least one PV during the first attempt. Additionally, 47.6% of operators reported they would have altered their strategy using 3D-EAM guidance. Overall lesion coverage was judged satisfactory in 77 of 84 PVs (91.5%). In the real-time 3D-EAM phase, no veins were missed. Compared with fluoroscopy-only and 3D-blinded procedures, 3D-EAM guidance significantly reduced the number of PVI applications (33.6 ± 7.8 vs. 38.4 ± 9.0 and 39.3 ± 10.5; p = 0.007) but increased left atrial dwell time (23.6 ± 7.2 vs. 16.7 ± 4.3 min; p < 0.001).
conclusionsReal-time 3D-EAM guidance improved PV targeting, reduced unnecessary applications, and eliminated missed veins. Integration of 3D mapping into PFA workflows is safe, feasible, and may enhance procedural quality. Larger studies are warranted to determine effects on long-term arrhythmia recurrence and durability.