ArticleJournal of applied clinical medical physics2026
Dynamic range optimization for treatment time reduction in respiratory-gated proton therapy using RayStation v2025.
Article in Journal of applied clinical medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Dynamic range optimization for treatment time reduction in respiratory-gated proton therapy using RayStation v2025.Journal of applied clinical medical physics · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
Abstract
purposeThis study introduces dynamic range optimization, which constrains the lower limit of the minimum MU per spot, as a planning-based strategy to enhance efficiency in respiratory-gated proton therapy and evaluates the resultant dosimetric trade-offs.
methodsWe analyzed 101 hepatocellular carcinoma patients who received line-scanning proton therapy. We developed a computational framework to predict the total treatment time, integrating dynamic range-mediated changes in beam-on time (BoT) with respiratory gating dynamics. The model was validated against fully dynamic range-optimized plans generated by RayStation v2025 across five dynamic range values (no constraint, 200, 100, 50, and 10). These values represent the ratio of maximum to minimum MU allowed within an energy layer; specifically, a lower dynamic range value imposes a stricter constraint by elevating the minimum MU floor, thereby reducing intensity modulation flexibility while increasing delivery speed. Dosimetric trade-offs were quantified using fully re-optimized plans for a representative case under dynamic range = 100 and 10 constraints. The study used a layer switching time (T
resultsDecreasing the dynamic range systematically reduced beam-on time in a sigmoid-shape pattern. The computational model's predictions for beam-on time showed a maximum error of 10.4% compared to RayStation v2025 calculations, confirming its accuracy. Total treatment time reduction reached a plateau at certain dynamic range thresholds, particularly when layer delivery plus switching was completed within a single gating-on period, indicating no further efficiency gain from additional dynamic range reduction. While target coverage remained equivalent across dynamic range values, lower dynamic range values systematically increased normal tissue doses due to reduced intensity modulation capability, with organs proximal to the target experiencing the largest relative increases (e.g., gallbladder mean dose increased 16.4% at DR = 10).
conclusionDynamic range optimization is an effective, planning-based method to reduce treatment time in respiratory-gated proton therapy, operating independently of patient cooperation or system modifications. The developed RayStation-integrated tool enables clinicians to identify the optimal patient-specific dynamic range, balancing efficiency gains against acceptable dosimetric quality, particularly for organs-at-risk adjacent to targets. This approach provides a technical foundation for individualized dynamic range selection during treatment planning.