In vitro and in vivo validation of a novel 3D-printed vessel anastomosis device for microvascular surgery.

John Ser Pheng Loh, Kuan-Che Feng, Yijia Yuan, Yinling Zhu, Robert Heymann, Lars Rasmusson, Justin Kok Soon Tan, Hwa Liang Leo, Reinhilde Jacobs

Abstract read
In one paragraph

Article in Scientific reports, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors.

John Ser Pheng Loh *Faculty of Dentistry, National University of Singapore (NUS), Singapore, Singapore. john.ser.pheng.loh@ki.se.
Kuan-Che Feng *Faculty of Dentistry, National University of Singapore (NUS), Singapore, Singapore.
Yijia Yuan *Faculty of Dentistry, National University of Singapore (NUS), Singapore, Singapore.
Yinling ZhuDepartment of Biomedical Engineering, College of Design and Engineering, National University of Singapore (NUS), Singapore, Singapore.
Robert HeymannDepartment of Dental Medicine, Karolinska Institutet, Stockholm, Sweden.
Lars RasmussonDepartment of Oral and Maxillofacial Surgery, Odontology Section 1, University of Gothenburg, Gothenburg, Sweden.
Justin Kok Soon TanDepartment of Biomedical Engineering, College of Design and Engineering, National University of Singapore (NUS), Singapore, Singapore.
Hwa Liang LeoDepartment of Biomedical Engineering, College of Design and Engineering, National University of Singapore (NUS), Singapore, Singapore.
Reinhilde JacobsDepartment of Dental Medicine, Karolinska Institutet, Stockholm, Sweden.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Microvascular anastomosis is fundamental to free-flap reconstruction, yet hand-sutured techniques remain highly skill-dependent and prolong ischemic time, contributing to thrombosis and flap loss. Current suture-less devices accelerate anastomosis but are constrained by vessel-size mismatch, eversion-related intimal injury, and inconsistent arterial performance. A clinically adaptable, arterial-capable system is needed. We developed a customizable, 3D-printed intraluminal coupler with a snap-fit connection and elastic external clasp that avoids vessel eversion and preserves length. Devices were fabricated via SLA or PolyJet printing using clinically used resins. Benchtop evaluation included burst-pressure testing, tensile testing, wettability and endothelial cytocompatibility assays with oxygen-plasma surface modification. Deployment was assessed ex vivo using porcine coronary vessels and in vivo in a porcine carotid arterial model with patency monitoring over 4 h. Couplers sustained leakage pressures of ~ 90 mmHg vs. ~16 mmHg for hand-sutured controls (P < 0.01), while maintaining comparable mechanical strength (≈ 2–3 N). Plasma surface treatment reduced water contact angles (≈ 85°→≈60°) and tripled endothelial attachment, restoring confluent morphology. Ex vivo deployment achieved completion in 9.47 ± 1.20 min, a ~ 62.5% reduction vs. published suturing times. In vivo, couplers restored immediate perfusion with no leakage or thrombosis. This 3D-printed intraluminal coupler demonstrates mechanical feasibility, rapid deploy-ability, and surface-modifiable endothelial compatibility in benchtop and short-term large-animal feasibility testing, supporting its potential for further preclinical development as a vascular anastomosis technology. Future survival studies and anti-thrombogenic surface engineering will advance readiness for clinical implementation.

Indexed as

Anastomosis, SurgicalMicrosurgeryPrinting, Three-DimensionalVascular Surgical ProceduresAnimalsCarotid ArteriesCoronary VesselsHumansSwine3D printedAnastomosisCouplerFree tissue flapMedical devicesMicrosurgery