Evidence map›Paper›PMID 41679927›Full record

ArticleJournal of applied clinical medical physics2026

The geometric accuracy of off-axis targets in stereotactic body radiotherapy treatments across three linear accelerators.

Dinesan Chinnaiya, Gopinath Mudhana

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors.

Dinesan ChinnaiyaDepartment of Physics, School of Advanced Sciences, Vellore Institute of Technology, Chennai, India.
Gopinath MudhanaDepartment of Physics, School of Advanced Sciences, Vellore Institute of Technology, Chennai, India.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThis study quantifies isocenter and off-axis geometric uncertainties using MultiMet Winston-Lutz (WL) tests to optimize gross tumour volume (GTV)-to-PTV (planning target volume) margin for single isocenter multiple target (SIMT) stereotactic body radiotherapy (SBRT) across three linear accelerators. METHODS AND MATERIALS: Geometric inaccuracies were quantified for Trilogy (HD-MLC, 6 MV SRS), TrueBeam (Millennium MLC), and TrueBeam STx (HD-MLC, 6-degrees-of-freedom (6DoF) couch) using a Sun Nuclear MultiMet-WL cube containing six tungsten carbide markers arranged along the superior-inferior axis. The electronic portal imaging device (EPID) images acquired at four cardinal gantry angles with varied collimator/couch rotations were analyzed using MultiMet-WL software (v2.1) to measure 3D (Δ) displacements for all the LINACs. The required GTV-to-PTV margins were calculated using a modified van Herk formula (2.5Σ+1.64σ), incorporating measured 3D displacements for isocenter and off-axis targets.

resultsThe TrueBeam STx (HD-MLC/6DoF) demonstrated superior geometric accuracy, maintaining ≤0.5 mm isocenter precision and ≤0.59 mm off-axis targeting (3-7 cm). The Trilogy exceeded TG-142 tolerances (1.06 ± 0.59 to 1.09 ± 0.57 mm) at all targets, requiring 4 mm uniform margins, while the TrueBeam (MMLC) showed optimal variations (isocenter: 0.68 ± 0.34 mm; superior off-axis: 0.74 ± 0.36 mm). Both TrueBeam platforms achieved sub-millimeter accuracy but demonstrated direction dependency for off-axis targets, requiring 2-3 mm anisotropic margins. Notably, isotropic margins introduced up to 11% delineation errors for off-axis targets due to these machine-specific geometric variations, highlighting the imperative for platform-specific margin protocols in SIMT SBRT.

conclusionThis study demonstrates that routine analysis of MultiMet WL testing is an essential tool for the spatial accuracy of the LINAC to establish machine-specific PTV margin expansion in SIMT-SBRT, particularly for targets where rotational errors dominate.

Indexed as

NeoplasmsParticle AcceleratorsPhantoms, ImagingRadiosurgeryRadiotherapy Planning, Computer-AssistedHumansRadiotherapy DosageRadiotherapy, Intensity-Modulatedgeometric accuracySIMT–SBRTWinston–Lutz test

Identifiers

PMID41679927
PMCPMC12900569

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.