Advanced preclinical testing of a design-optimized ceramic bearing for a cervical total disc replacement.

Lucia Kölle, Gregory Pryce, Andrew R Beadling, Michael Bryant, Richard M Hall, Stephen J Ferguson, Benedikt Helgason

Abstract read
In one paragraph

Article in PloS one, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Lucia KölleDepartment of Health Sciences and Technology, ETH Zürich, Zürich, Switzerland.ORCID https://orcid.org/0000-0003-3969-2772
Gregory PryceInstitute for Functional Surfaces, School of Mechanical Engineering, University of Leeds, Leeds, United Kingdom.
Andrew R BeadlingDepartment of Mechanical Engineering, School of Engineering, University of Birmingham, Birmingham, United Kingdom.
Michael BryantDepartment of Mechanical Engineering, School of Engineering, University of Birmingham, Birmingham, United Kingdom.
Richard M HallDepartment of Mechanical Engineering, School of Engineering, University of Birmingham, Birmingham, United Kingdom.ORCID https://orcid.org/0000-0001-5504-6717
Stephen J FergusonDepartment of Health Sciences and Technology, ETH Zürich, Zürich, Switzerland.
Benedikt HelgasonDepartment of Health Sciences and Technology, ETH Zürich, Zürich, Switzerland.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectiveArthroplasty in the cervical spine is performed using Total Disc Replacement (TDR) implants. This study investigates the safety and tribology of a design-optimized ceramic TDR bearing.

methodsSamples of an optimized bearing geometry were manufactured from the bioceramic BIOLOX®delta (CeramTec GmbH, Germany) and tested in a six degree of freedom universal joint simulator in diluted calf serum at body temperature. The investigated load and motion profiles simulated movements of daily living; three pure rotations under static load and three rotations coupled under sinusoidal load as well as adverse events (three subluxation test conditions).

resultsDynamic coefficients of friction during flexion/extension and lateral bending were µ = 0.16 ± 0.011 and µ = 0.16 ± 0.070, respectively. Friction factors during flexion/extension and lateral bending were f = 0.14 ± 0.0092 and f = 0.13 ± 0.063, whereas the friction factor in axial rotation was f = 0.033 ± 0.0028 and for three coupled rotations under sinusoidal load, it was f = 0.13 ± 0.028. The design demonstrated a sufficient resistance against subluxation, with subluxation forces more than 3.5-fold the 20N that are typically required as acceptance criteria in FDA safety and effectiveness data.

conclusionFriction values were found to be comparable to or higher than those typically reported for ceramic-on-ceramic hip replacements. Furthermore, the results indicated that the optimized bearing design is sufficiently resistant to subluxation. Significance: Test protocols to investigate safety and tribology of cervical TDRs are proposed and their results are reported for a design-optimized ceramic TDR bearing.

Indexed as

CeramicsCervical VertebraeProsthesis DesignTotal Disc ReplacementAnimalsFrictionHumansMaterials TestingWeight-Bearing