Does the bone drill bit remain sharp after a collision with the intramedullary pin? a study on isolated canine tibias.
Article in BMC veterinary research, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundBlunting of the tip, as well as smoothing and abrasion of the cutting edges, are natural signs of wear observed under magnification after repeated use of bone drill bits. The lack of a standardized surgical protocol for using drill bits in human and veterinary surgery often results in their use beyond the manufacturer’s recommended limits. Drilling with a dull drill bit results in excessive local material degradation, increased surface roughness, and inaccurate hole dimensions. In bone surgery, an additional factor is thermal damage to the bone. Colliding with a metallic object while rotating can cause the cutting edges of the drill bit to become blunt or chipped, or the drill bit to bend or break. This study aimed to analyze damage to bone drill bits resulting from collisions with an intramedullary Kirschner wire.
methodsThe experiment was performed on the left and right isolated tibias of an adult female Boxer dog. Kirschner wires were inserted intramedullary, and holes were drilled in mediolateral and craniocaudal directions to mimic plate-rod osteosynthesis. The drill bits were evaluated for suitability for repeated use based on an examination using a binocular stereomicroscope and digital microscope.
resultsOne bone drill bit broke. Only 17 / 44 unbroken bone drill bits (39%) could be reused following collision with the Kirschner wire. Twenty-seven drill bits sustained significant damage to their main cutting edges and were deemed unsuitable for reuse.
conclusionsContact between the rotating drill bit and the intramedullary implant is undesirable. If this occurs, the drill bit should be examined under a microscope to determine its suitability for further use.