Comparative Dosimetry of Handheld Versus Wall-Mounted Intraoral X-ray Systems: A Phantom-Based Study.

Vignesh Guptha Raju, Rajsandeep Singh, Ameer Akhil Ahmed Shaik, Sai Priyanka Gaddipati, Venkat Hemant Akurati, Niti Dharmendra Shah, Rahul Vc Tiwari, Seema Gupta

Abstract read
In one paragraph

Article in Cureus, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors.

Vignesh Guptha RajuDepartment of Pedodontics and Preventive Dentistry, Karpaga Vinayaga Institute of Dental Sciences, Chennai, IND.
Rajsandeep SinghDepartment of Dentistry, Punjab Civil Secretariat, Chandigarh, IND.
Ameer Akhil Ahmed ShaikDepartment of Oral Medicine and Radiology, Narayana Dental College and Hospital, Nellore, IND.
Sai Priyanka GaddipatiDepartment of Dentistry, P.M. Nadagouda Memorial Dental College and Hospital, Bagalkot, IND.
Venkat Hemant AkuratiDepartment of General Dentistry, School of Dentistry, University of Louisville, Louisville, USA.
Niti Dharmendra ShahDepartment of Global Management, Thunderbird School of Global Management, Phoenix, USA.
Rahul Vc TiwariDepartment of Oral and Maxillofacial Surgery, RKDF Dental College and Research Center, Bhopal, IND.
Seema GuptaDepartment of Orthodontics, Kothiwal Dental College and Research Centre, Moradabad, IND.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

introductionThe rapid adoption of handheld dental X-ray units has revolutionized imaging in mobility-restricted patients; however, concerns may still persist regarding their radiation safety profile compared with traditional wall-mounted systems. The aim of the present study was to quantitatively compare operator- and patient-equivalent radiation doses, ambient scatter, and diagnostic image quality between a handheld (MaxRay Cocoon, Dexcowin Co. Ltd., Korea) and a wall-mounted (Optima DC, Alerio Technologies Pvt. Ltd., India) intraoral X-ray unit during standardized mandibular molar radiography. MATERIALS AND

methodsThis prospective experimental study was conducted using a human cadaver mandible as a standardized phantom. A total of 384 exposures (198 per device) were performed with identical exposure time (0.18 seconds), rectangular collimation, and a size-2 complementary metal-oxide-semiconductor (CMOS) digital sensor using the paralleling technique. Operator doses were recorded using real-time electronic personal dosimeters (RaySafe i3, RaySafe, Sweden) at the chest and thyroid levels and thermoluminescent dosimeters (TLD-100) placed under the lead apron at the chest, forehead, and fingers of the dominant hand. The patient-equivalent effective dose was calculated from dose-area product (DAP) measurements using conversion coefficients recommended by the International Commission on Radiological Protection. Image quality was evaluated objectively through signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) measurements and subjectively by three blinded experienced oral radiologists using a five-point Likert scale. Data were analyzed using independent sample t-tests.

resultsThe handheld unit produced significantly higher operator doses: finger dose 0.45 ± 0.12 µSv vs. 0.08 ± 0.03 µSv (p < 0.001), chest (outside apron) 0.18 ± 0.05 µSv vs. 0.05 ± 0.02 µSv (p = 0.001), and thyroid 0.11 ± 0.03 µSv vs. 0.04 ± 0.01 µSv (p = 0.003). The effective dose to the mandible phantom was 2.65 ± 0.25 µSv (handheld) vs. 1.82 ± 0.18 µSv (wall-mounted) (p < 0.001). The ambient scatter at 1 m was three times higher for the handheld device (p < 0.001). No significant differences were found in SNR (p = 0.125), CNR (p = 0.08), or subjective diagnostic quality scores (p = 0.65).

conclusionsDespite comparable diagnostic image quality, the handheld dental X-ray unit delivered substantially higher radiation doses to both the operator and the patient-equivalent phantom than the wall-mounted system. Wall-mounted units should remain the standard where patient mobility permits, reserving handheld systems for clinically justified scenarios.

Indexed as

dosimetryexposureoperatorradiationsafety

Identifiers

PMID41589187
PMCPMC12831817

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials