Evidence map›Paper›PMID 41583941›Full record

ArticleJournal of experimental orthopaedics2026

Reliability of a wireless instrumented insole (WalkinSense system) for measuring spatiotemporal and kinematic gait variables.

Melanie Eckelt, Jennifer Fayad, Anne Backes, Frederic Garcia, Bernd Grimm, Valeria Serchi, Tobias Meyer, Thomas Solignac, Caroline Mouton, Romain Seil and 1 more

Abstract read
In one paragraph

Article in Journal of experimental orthopaedics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

11 authors.

Melanie EckeltDepartment of Precision Health Luxembourg Institute of Health Strassen Luxembourg.ORCID https://orcid.org/0000-0002-5835-1116
Jennifer FayadDepartment of Precision Health Luxembourg Institute of Health Strassen Luxembourg.ORCID https://orcid.org/0000-0001-6665-0696
Anne BackesDepartment of Precision Health Luxembourg Institute of Health Strassen Luxembourg.ORCID https://orcid.org/0000-0003-2026-5762
Frederic GarciaDepartment of Precision Health Luxembourg Institute of Health Strassen Luxembourg.ORCID https://orcid.org/0000-0001-7623-3175
Bernd GrimmDepartment of Precision Health Luxembourg Institute of Health Strassen Luxembourg.ORCID https://orcid.org/0000-0003-4585-4643
Valeria SerchiIEE S.A. Bissen Luxembourg.ORCID https://orcid.org/0000-0002-1661-0060
Tobias MeyerIEE S.A. Bissen Luxembourg.ORCID https://orcid.org/0009-0004-9569-3029
Thomas SolignacIEE S.A. Bissen Luxembourg.
Caroline MoutonDepartment of Orthopaedic Surgery Centre Hospitalier Luxembourg-Clinique d'Eich Eich Luxembourg.ORCID https://orcid.org/0000-0002-9582-6870
Romain SeilDepartment of Orthopaedic Surgery Centre Hospitalier Luxembourg-Clinique d'Eich Eich Luxembourg.ORCID https://orcid.org/0000-0001-8806-9384
Laurent MalisouxDepartment of Precision Health Luxembourg Institute of Health Strassen Luxembourg.ORCID https://orcid.org/0000-0002-6601-5630

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Purpose: Reliable gait analysis is essential for clinical assessment and research. Wearable technologies such as the WalkinSense system (WSS), a wireless instrumented insole system equipped with an inertial measurement unit, enable the measurement of spatiotemporal and kinematic gait parameters in real-world settings. The accuracy of the WSS has been previously validated. This study aims to evaluate the test-retest reliability of the WSS in healthy adults walking and running under various speed and slope conditions. Methods: Forty-nine healthy adults completed two sessions one week apart, walking and running at 3, 4.5 (-3°, -6°, +3°, +6°), 6 and 9 km/h on a treadmill. Spatiotemporal variables, including stance time, swing time, stride time, stride length, single and double support time, as well as kinematic variables such as foot ground angle at initial contact and toe-off, were recorded using the WSS. Relative reliability was assessed using intraclass correlation coefficients (ICC Results: Results demonstrated good to excellent reliability for most spatiotemporal parameters across all conditions, with ICC values ranging from 0.76-0.95, while the foot ground angles exhibited lower reliability (ICC: 0.71-0.86). SEM% and MDC% were generally below 10% for spatiotemporal measures (SEM%: 1.63-6.63; MDC%: 3.44-18.36), reflecting both low measurement error and high sensitivity to detect real changes beyond random variation. Bland-Altman analyses revealed no relevant heteroscedasticity. Conclusion: These findings support the WSS as a reliable tool for assessing spatiotemporal variables in healthy adults across diverse walking and running conditions, underscoring its potential for use in both clinical and research environments. However, clinical studies are needed to fully establish its utility in patient assessment. Level of Evidence: Level II, diagnostic studies.

Indexed as

gait analysisinstrumented insolestest–retest reliabilitywearable sensors

Identifiers

PMID41583941
PMCPMC12825025

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.