Agreement and repeatability of intraocular pressure measurements between SK-5500B and Icare IC200 tonometers in sitting and supine positions: a comparative study.

Nan Li, Yucong Chen, Kunliang Qiu, Zhiqiang Guan

Abstract read
In one paragraph

Article in Frontiers in physiology, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Nan LiJoint Shantou International Eye Center of Shantou University and The Chinese University of Hong Kong, Shantou, China.
Yucong ChenJoint Shantou International Eye Center of Shantou University and The Chinese University of Hong Kong, Shantou, China.
Kunliang Qiu *Joint Shantou International Eye Center of Shantou University and The Chinese University of Hong Kong, Shantou, China.
Zhiqiang Guan *Joint Shantou International Eye Center of Shantou University and The Chinese University of Hong Kong, Shantou, China.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Introduction: Accurate intraocular pressure (IOP) measurement is vital in glaucoma management, yet the gold standard (Goldmann applanation tonometry) is limited to seated measurements. Rebound tonometers like the Icare IC200 and the newer SK-5500B offer portability for assessing posture-induced IOP changes, but their measurement consistency requires validation. This study aims to evaluate the agreement and repeatability of IOP measurements between the SK-5500B and Icare IC200 tonometers in sitting and supine positions. Methods: This prospective study recruited 59 subjects (97 eyes). IOP was measured in both sitting and supine positions by two masked operators using the SK-5500B and Icare IC200 tonometers in a randomized sequence. Subjects were categorized into four groups based on their baseline non-contact tonometry (NCT) readings (<10 mmHg, 10-21 mmHg, 22-30 mmHg, and >30 mmHg). Agreement was assessed using Pearson's correlation and Bland-Altman analysis. Correlation was evaluated using Pearson's correlation coefficient (r). Intra- and inter-operator repeatability were determined by within-subject standard deviation (Sw), coefficient of variation (CV), and intraclass correlation coefficient (ICC). The correlation between the device and central corneal thickness (CCT) or axial length (AL) was also analyzed. Results: The SK-5500B and Icare IC200 showed a strong positive correlation across all conditions (both positions and both operators) (r ≥ 0.90, Conclusion: Despite wide limits of agreement in Bland-Altman analysis, the SK-5500B showed good agreement and highly comparable repeatability with the Icare IC200 in both sitting and supine positions, supporting its potential as a practical clinical alternative.

Indexed as

agreementIcare ic200intraocular pressurerebound tonometryrepeatabilitySK-5500B

Identifiers

PMID41574191
PMCPMC12823037

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.