Evidence map›Paper›PMID 41574155›Full record

ArticleJournal of wrist surgery2026

Failure of Universal 2 Wrist Joint Replacements: A Retrieval Study.

Thomas J Joyce, Göksu Kandemir, Nasri H Zreik, Sumedh Talwalkar, Ian Trail

Abstract read
In one paragraph

Article in Journal of wrist surgery, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Measurement of the wear volumes generated within retrieved magnetically controlled growing rods.Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine · 2026
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Thomas J JoyceSchool of Engineering, Newcastle University, Newcastle upon Tyne, United Kingdom.ORCID 0000-0002-8497-5790
Göksu KandemirSchool of Engineering, Newcastle University, Newcastle upon Tyne, United Kingdom.
Nasri H ZreikDepartment of Upper Limb Surgery, Wrightington Hospital, Wigan, United Kingdom.
Sumedh TalwalkarDepartment of Upper Limb Surgery, Wrightington Hospital, Wigan, United Kingdom.
Ian TrailDepartment of Upper Limb Surgery, Wrightington Hospital, Wigan, United Kingdom.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: The Universal 2 total wrist arthroplasty was one of the most common wrist replacements, but long-term results were disappointing, due to substantial damage to the polyethylene component and, in some cases, metallosis. The purpose of this research was to investigate the underlying reasons for this polyethylene damage. Materials and Methods: From a single clinical center, retrieval analysis was undertaken on six Universal 2 wrists and two additional polyethylene components. All components were analyzed at the macroscale and dimensional measurements of polyethylene components were undertaken. These were compared against known component orientation in vivo to identify areas of greatest material loss. Results: Of 62 Universal 2 wrists implanted, there was a tendency for smaller implants to be revised more frequently. Of the six explanted wrists, material loss was always noticeable on the ulnar side, and to a lesser extent in the dorsal direction. Five of eight polyethylene components had failed at the base of the ulnar-side blind hole. Smaller implants tended to have less thickness at the base of the blind holes, thus explaining the failure of smaller sizes. Discoloration and surface fatigue of explants indicated oxidation of components. Average time in vivo for the explants was 13.8 years which indicates the slow nature of polyethylene oxidation. All were revised due to loosening. All were implanted in rheumatoid patients with a mean age at surgery of 56.1 years. Three of the six Universal 2 wrists, all sized Extra Small, showed severe wear of the titanium alloy carpal component. Conclusion: Failure of the polyethylene components was due to a combination of inappropriate sterilization technique plus a design issue where polyethylene was thinnest on the smallest size components. Continued surveillance of patients implanted with Universal 2 wrists is recommended.

Indexed as

failureretrieval analysiswearwrist arthroplastywrist replacement

Identifiers

PMID41574155
PMCPMC12823192

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.