Image-Guided Microwave Ablation of Liver Tumours Using a 150-W Generator: Ex Vivo and Multicenter in Vivo Predictability Assessment.
Article in Cardiovascular and interventional radiology, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
13 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo assess the predictability of Emprint HP-Generator (150 W, 2.45 GHz) in terms of ablation volume and sphericity both in the ex vivo and in vivo setting. MATERIALS AND
methodsIn an ex vivo setting, 48 ablations were performed on porcine livers at varying ablation times with a constant power of 150 W. MRI assessment of the ablation volumes was performed. In the in vivo setting consecutive procedures with 150W application with no rampage, single antenna positioning and availability of a contrast-enhanced computed tomography (CECT) immediately after the ablation were included. A semi-automated segmentation tool was used for volume estimation and spherical index (SI) was calculated.
resultsIn the ex vivo group, mean ablation time was 5.3 ± 2.5 min (range 1-10 min), mean ablation diameter was 37.7 ± 5.8 mm (range 21.3-49 mm) and mean ablation volume was 20.1 ± 10.8 cm
conclusionThe Emprint HP-generator enables the creation of large predictable volumes in both settings. SI remains consistent regardless of ablation time and volume.