Analytical Performance of Two Flow Cytometry Systems for CD34⁺ Hematopoietic Stem Cell and CD3⁺ T-Cell Enumeration in Stem Cell Transplantation.

Habsah Aziz, Suria Abdul Aziz

Abstract read
In one paragraph

Article in Cureus, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors.

Habsah AzizInstitute of Systems Biology, National University of Malaysia, Bangi, MYS.
Suria Abdul AzizDepartment of Pathology, Faculty of Medicine, Universiti Kebangsaan Malaysia, Kuala Lumpur, MYS.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background Accurate enumeration of hematopoietic stem cells (HSC) and T-lymphocytes is essential for stem cell transplantation (SCT), as engraftment success depends on the infusion of adequate CD34⁺ and CD3⁺ cell doses. Flow cytometry is the gold standard for quantifying these subsets; however, analytical performance may vary between platforms, necessitating formal validation before clinical implementation. Therefore, this study aimed to evaluate the analytical precision of the DxFLEX flow cytometer and to assess its correlation and agreement with the established FACSCalibur reference method for CD34⁺ HSC and CD3⁺ T-cell enumeration. Materials and methods Analytical precision and method-comparison studies were conducted in accordance with Clinical and Laboratory Standards Institute (CLSI) EP15-A3 guidelines. Two levels of third-party quality control (QC) materials were used for CD34⁺ and CD3⁺ assays. Within-run precision was assessed using ten replicates per QC level (n =10), while between-run precision was evaluated across five days with three runs per day (n = 15). Mean values, standard deviations (SD), and coefficients of variation (CV) were compared with QC performance specifications. Method correlation was performed using 40 specimens following CLSI H57-A guidelines. Correlation coefficients and Bland-Altman analyses were used to assess agreement between the evaluated system (DxFLEX) and the reference analyzer (FACSCalibur). Results Both within-run and between-run precision for CD34⁺ and CD3⁺ measurements fell within acceptable QC limits, demonstrating stable assay performance. Correlation analyses showed excellent agreement between the two flow cytometry platforms, with correlation coefficients (r = 0.999 for CD34⁺ absolute counts and r = 1.00 for CD3⁺ absolute counts). Agreement between instruments was further supported by Bland-Altman analysis. For CD34⁺ measurements, the relative mean bias was -1.2% (limits of agreement: -33.4% to 31.1%) for percentage values and -0.479 (limits of agreement: -20.415 to 19.457) for absolute counts. For CD3⁺ measurements, the mean bias was -0.3% (limits of agreement: -3.5% to 2.9%) for percentage values and -0.358 (limits of agreement: -14.129 to 13.413) for absolute counts. These findings indicate strong concordance between the evaluated system and the reference platform, with minimal systematic differences and clinically acceptable limits of agreement. Conclusion The evaluated flow cytometry platform demonstrated excellent analytical performance for CD34⁺ and CD3⁺ enumeration, with precision and correlation results meeting clinical laboratory standards. Strong agreement with the reference method supports its suitability for routine use in HSC monitoring and SCT-related immunophenotyping.

Indexed as

cd34⁺ hematopoietic stem cellscd3⁺ t-cellcorrelationflow cytometryprecisionstem cell transplant