Validation of the hanvon KSY8600 blood pressure monitor in the general population based on ISO 81060-2:2018 + AMD1:2020 protocol.
Article in Journal of human hypertension, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
This study evaluated the accuracy of the Hanvon KSY8600 blood pressure monitor in adult populations in accordance with the AAMI/ESH/ISO (ISO 81060-2:2018) and AMD 1:2020 standards. 90 people were recruited as research participants, among which 88 eligible participants were chosen and analyzed according to the research plan. The average age of these participants was 51.3 ± 17.35 years old, with an average arm circumference of 28.1 ± 3.41 cm, and with females accounting for 60.2%. The test results showed that the mean differences (standard deviations) in systolic blood pressure (SBP) and diastolic blood pressure (DBP) displayed on the blood pressure monitor were -0.4 mmHg (2.09 mmHg) and -0.3 mmHg (1.86 mmHg) respectively, meeting the requirements of Criterion 1 (≤ ± 5 mmHg (8 mmHg)). Additionally, the average standard deviations of the participants' SBP and DBP were 0.91 mmHg and 0.90 mmHg respectively. According to Table 1 of Criterion 2, the mean values of SBP and DBP were -0.4 mmHg and -0.3 mmHg respectively, with maximum allowable standard deviations of SBP and DBP of ≤6.93 and ≤6.95 mmHg respectively, meeting the requirements of Criterion 2. It was thus proved that the blood pressure monitor complies with ISO 81060-2:2018 + AMD1:2020, and is recommended for clinical and home blood pressure measurement for adults.