Assessing imaging performance of ultrasound systems using a random hypoechoic sphere phantom with freehand scanning.
Article in Medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Assessing imaging performance of ultrasound systems using a random hypoechoic sphere phantom with freehand scanning.Medical physics · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
Abstract
backgroundDiagnostic ultrasound is rapidly evolving, and the increasing complexity of ultrasound systems underscores the importance of robust quality assurance (QA) and quality control (QC) methods. Current methods typically rely on manual image acquisition and subjective evaluation, making them operator-dependent, poorly reproducible, and challenging for longitudinal tracking. Efforts have been made to develop automated ultrasound QA/QC methods for more quantitative and reproducible outcomes. One such approach utilized the Random Hypoechoic Sphere Phantom (RHSP) to evaluate the detectability of ultrasound systems by measuring the human observer-related Lesion Signal-to-Noise Ratio (LSNR). A practical manual scanning method was developed for the RHSP, which eliminated the need for a mechanical guide to control transducer translation. However, this approach requires maintaining a uniform translation speed and a small translation distance between successive images. Broader adoption of the RHSP also requires easily accessible automated analysis software. PURPOSE: To improve the practicality of using the RHSP for ultrasound performance evaluation in clinical settings, we developed an automated analysis method tailored for freehand scanning. We validated the method by evaluating ultrasound system performance under various acquisition settings, identifying sources of variability, and analyzing their impact on LSNR measurements.
methodsThe RHSP consisted of 2 mm diameter spheres at 20% volume fraction. It was scanned by moving the transducer by freehand during a cine-loop acquisition. Volumetric images were generated by stacking the elevational frames. After denoising and homogeneity correction, the spheres were segmented in 3D using depth-adaptive thresholds from a Gaussian-Mixture Model. LSNR was computed for each segmented sphere, and LSNRs at similar depth were averaged to generate an LSNR vs. depth curve. The algorithm was validated by computing the LSNR vs. depth and sphere-count vs. depth curves for various acquisition settings, with the following factors changing one at a time: nominal frequency, imaging modes and compound techniques, transmit power, gain, dynamic range, and transmit focal depth. Variability of the algorithm was assessed using repeated scans performed by different operators and at various transducer translation speeds.
resultsAll LSNR results closely matched visual assessments and followed expectations when changing acquisition settings. LSNR values improved with higher nominal frequency, harmonic imaging with compound, higher transmit power, and decreased dynamic range; gain had minimal effect. The best LSNR also aligned with the transmit focal depths set by the operator. The variabilities of LSNR were from 2.5% to 6.9% intra-operator, 5.2% inter-operator, and 5.5% due to transducer translation speed. These variabilities of the LSNR were much smaller than those of the corresponding sphere counts, demonstrating the proposed algorithms' robustness against potential variations produced by freehand scanning.
conclusionsThe automated analysis method for the RHSP with freehand scanning provides an accurate and stable quantitative assessment of lesion detectability for evaluating ultrasound performance, and is feasible in clinical settings. It demonstrated robustness by achieving approximately 5% variability in LSNR across various transducer translation speeds and different operators. With such robustness, this approach simplifies ultrasound performance evaluation and holds promise for integration into routine ultrasound QA/QC in a clinical environment.