Evidence map›Paper›PMID 41508105›Full record

ArticleJournal of orthopaedic surgery and research2026

Validation and reliability of a novel portable automated knee arthrometer: a cadaveric study bridging the skin-to-bone gap and suggesting medial pivotal rotation as a superior marker for ACL deficiency.

Geunwu Gimm, Sungwoo Park, Man Soo Kim, Hyuk-Soo Han, Sungwan Kim

Abstract readValidation Study
In one paragraph

Article in Journal of orthopaedic surgery and research, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Geunwu Gimm *Department of Biomedical Engineering, Seoul National University College of Medicine, Seoul, 03080, Republic of Korea.ORCID http://orcid.org/0000-0002-7414-4125
Sungwoo Park *Institute of Convergence Medicine with Innovative Technology, Seoul National University Hospital, Seoul, 03122, Republic of Korea.ORCID http://orcid.org/0000-0002-0027-5085
Man Soo KimDepartment of Orthopaedic Surgery, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, 06591, Republic of Korea.ORCID http://orcid.org/0000-0002-1374-9964
Hyuk-Soo HanDepartment of Orthopaedic Surgery, Seoul National University Hospital, Seoul, 03080, Republic of Korea. oshawks7@snu.ac.kr.ORCID http://orcid.org/0000-0003-1229-8863
Sungwan KimDepartment of Biomedical Engineering, Seoul National University College of Medicine, Seoul, 03080, Republic of Korea. sungwan@snu.ac.kr.ORCID http://orcid.org/0000-0002-9318-849X

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundAccurate measurement of tibial rotation in anterior cruciate ligament (ACL) injury remains challenging because most devices rely on skin-referenced measurements and do not account for concurrent femoral rotation. Moreover, conventional assessments focus on central axial internal rotation (CAIR) rather than the more physiological medial pivotal rotation (MPR), which reflects the pivot-shift phenomenon. This study aimed to (1) develop a portable automated knee arthrometer to measure anterior translation (AT), CAIR, and MPR; (2) assess its reliability and validity in skin- and bone-referenced measurements; (3) investigate the skin-bone relationship considering concurrent femoral rotation; and (4) compare MPR with conventional CAIR for clinical utility.

methodsSix cadaveric lower limbs were evaluated under intact, ACL-sectioned, and combined ACL- and anterolateral ligament-sectioned conditions. AT was measured using a laser displacement sensor. Tibial and femoral rotations were measured using skin- and bone-referenced sensors. Intrarater and interrater reliability were assessed, and validity was evaluated by comparison with Lachmeter for AT and Mako for CAIR. The relationship between skin- and bone-referenced rotation was analyzed using regression, and the increase in rotation after ACL sectioning was compared between MPR and CAIR.

resultsThe arthrometer demonstrated excellent intrarater and interrater reliability for AT, CAIR, and MPR (ICCs > 0.90). It showed higher interrater reliability than the Lachmeter in injured knees (P = .004). Validity against reference devices was acceptable. Although skin- and bone-referenced rotational measurements differed significantly (P < .001), cadaver-wise regression showed strong correlation for net tibial rotation (R²: 0.833 for CAIR; 0.715 for MPR). The increase in MPR after ACL sectioning was significantly greater than that of CAIR (4.31° vs. 1.49° at 4 N·m, P < .001).

conclusionsThe arthrometer showed excellent reliability and acceptable validity for assessing AT, CAIR, and MPR. Although skin-referenced measurements differ from true osseous rotation, they may provide clinically meaningful assessments for within-subject comparisons. MPR increased more than CAIR after ACL sectioning, suggesting it may serve as a superior marker for ACL deficiency.

Indexed as

Anterior Cruciate Ligament InjuriesArthrometry, ArticularKnee JointAnterior Cruciate LigamentCadaverFemaleHumansMaleRange of Motion, ArticularReproducibility of ResultsRotationTibiaAnterior translationCentral axial internal rotationMedial pivotal rotationPortable automated knee arthrometerSkin/bone-referenced rotational measurements.

Identifiers

PMID41508105
PMCPMC12882578

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.