ArticlePediatric radiology2026
Effect of shortening the fluoroscopy time audible alarm on total fluoroscopy time and reference air kerma for pediatric modified barium swallow studies.
Article in Pediatric radiology, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundFederal regulations require an audible alarm for each passage of 5 min of cumulative x-ray irradiation time during a diagnostic fluoroscopy procedure. In contrast to United States federal requirements, regulations vary amongst individual states.
objectiveDetermine whether a shortened fluoroscopy time alarm reduces pediatric patient ionizing radiation exposure for modified barium swallow studies (MBSS). MATERIALS AND
methodsFluoroscopy time alarms were changed from the passage of every 5 min to 2 min. A retrospective analysis of MBSS compared patient reference point air kerma and fluoroscopy time before and after the alarm time was shortened.
resultsA total of 3,875 MBSS performed on patients under the age of 18 years were analyzed. Reductions of 14.9% and 19.6% for average and median reference point air kerma (P<0.001) and 15.5% and 16.2% for average and median fluoroscopy time (P<0.001) were observed at the main campus for MBSS performed before and after changing the fluoroscopy time alarm from 5 min to 2 min. Reductions of 14.3% and 9.1% for average and median reference point air kerma (P<0.001) and 13.9% and 8.3% for average and median fluoroscopy time (P<0.001) were observed at the satellite locations for MBSS performed before and after changing the fluoroscopy time alarm.
conclusionThis retrospective study suggests a shortened fluoroscopy time alarm can significantly lower pediatric patient dose and fluoroscopy time in MBSS. State and federal regulations should consider establishing flexible fluoroscopic alarm settings, especially when used for the pediatric population.