Evidence map›Paper›PMID 41503054›Full record

ArticlePractical laboratory medicine2026

Assessment of the stability of 20 biochemical analytes in serum and whole blood samples after storage at nonstandard temperatures.

Gabriella Iacovetti, Bradley B Collier, Jill M Rafalko, Mitchell Peevler, Nicolas Tokunaga, Jason Ragar, Whitney C Brandon, Matthew R Chappell, Russell P Grant, Greg J Sommer and 1 more

Abstract read
In one paragraph

Article in Practical laboratory medicine, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

11 authors.

Gabriella IacovettiDiagnostic Devices Research and Development, Labcorp, 6624 Owens Dr., Pleasanton, CA, 94588, United States.
Bradley B CollierResearch & Development, Labcorp, 1447 York Court, Burlington, NC, 27215, United States.
Jill M RafalkoResearch & Development, Labcorp, 1447 York Court, Burlington, NC, 27215, United States.
Mitchell PeevlerDiagnostic Devices Research and Development, Labcorp, 6624 Owens Dr., Pleasanton, CA, 94588, United States.
Nicolas TokunagaDiagnostic Devices Research and Development, Labcorp, 6624 Owens Dr., Pleasanton, CA, 94588, United States.
Jason RagarDiagnostic Devices Research and Development, Labcorp, 6624 Owens Dr., Pleasanton, CA, 94588, United States.
Whitney C BrandonResearch & Development, Labcorp, 1447 York Court, Burlington, NC, 27215, United States.
Matthew R ChappellResearch & Development, Labcorp, 1447 York Court, Burlington, NC, 27215, United States.
Russell P GrantResearch & Development, Labcorp, 1447 York Court, Burlington, NC, 27215, United States.
Greg J SommerDiagnostic Devices Research and Development, Labcorp, 6624 Owens Dr., Pleasanton, CA, 94588, United States.
Ulrich Y SchaffDiagnostic Devices Research and Development, Labcorp, 6624 Owens Dr., Pleasanton, CA, 94588, United States.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: Blood specimen transport conditions can have critical impacts on analyte stability and test result accuracy. This study evaluated the impact of multiday storage of specimens at 30 °C and 0 °C, approximating shipment under summer conditions and shipment in contact with melting ice, respectively. Methods: Blood samples from 16 healthy subjects were processed as serum samples and as clotted whole blood (uncentrifuged) stored at 30 °C on separator gel, and as serum stored in microtainers at 0 °C, each for 24- and 72-h periods prior to analysis. Each sample was analyzed for 20 common analytes and compared to the values from a paired baseline control sample. Mean absolute and/or relative biases were calculated and compared to CLIA acceptance limits. Results: Serum samples stored at 30 °C for 24- or 72-h met acceptance criteria for all assessed analytes, except for 72-h ALT, Total Bilirubin, Carbon Dioxide, Creatinine, and Sodium. Numerous analytes were unstable at 30 °C in uncentrifuged whole blood, with only Albumin, ALP, Total Bilirubin, Cholesterol, HDL, Total Protein, Triglycerides, Uric Acid, and hsCRP remaining stable up to 72 h. For serum samples stored at 0 °C, all 24- hour and 72-h analytes showed biases within limits. Conclusion: Storage of serum samples at 0 °C for up to 72 h yielded valid results for all analytes studied. Storage at 30 °C for up to 72-h yielded valid results for most analytes when combined with centrifugation prior to storage. Compared to room temperature storage, ALT and Sodium in serum and whole blood showed signs of accelerated degradation at 30 °C.

Indexed as

Analyte stabilityCentrifugationClinical validationCold chainPre-analytical errorSpecimen transportStability testing

Identifiers

PMID41503054
PMCPMC12769415

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.