ArticleEuropean archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery2026
Titanium incus interposition in ossiculoplasty: A retrospective study.
Article in European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeThis study aimed to evaluate the outcomes of ossiculoplasty using titanium incus interposition prostheses, focusing on audiological benefits, extrusion and dislocation rates.
methodsThis retrospective cohort study was conducted at a tertiary referral center and included 27 patients with conductive hearing loss who underwent ossiculoplasty using titanium incus interposition prostheses between 2015 and 2023. Preoperative and postoperative (minimum 6 months) otoscopic examinations and audiological assessments (average thresholds at 0.5, 1, 2, and 3 kHz) were performed. Patients were categorized as "stable" (n = 20, 74%) if the prosthesis remained functional, or as "failure" (n = 7, 26%) if reoperation was required for middle ear complications.
resultsThe mean postoperative air bone gaps (ABG) across 0.5-3 kHz were 19.6 (9.3) dB for all patients, and 16.1 (7.6) dB in the stable group. A total of 20 patients (74%) achieved either a postoperative ABG less than 20 dB or air conduction improvement more than 15 dB. Prosthesis dislocation occurred in 3 cases (11%), with a mean time of 31 months (range: 14-61 months). Extrusion was observed in 3 cases (11%), with a mean time of 25 months (range: 11-34 months). No significant difference in postoperative ABG or ABG closure was observed with respect to surgical indication, surgical approach, or the degree of temporal bone pneumatization.
conclusionsTitanium incus interposition prostheses can provide satisfactory postoperative hearing in patients with an intact stapes, preserved malleus handle, and favorable ossicular alignment. Nevertheless, prosthesis displacement or extrusion occurred in 22% of cases, highlighting the need for improved design, biomaterials, and placement techniques to enhance long-term stability.