ArticleMagnetic resonance in medicine2026
Rapid Free-Breathing and Automated 2D Shimming of the Lung at 3T.
Article in Magnetic resonance in medicine, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
Abstract
purposeTo develop and validate a rapid, fully automated method for shimming of the lung.
methodsFor field mapping of the lung, a custom 2D RF spoiled gradient echo sequence was used offering sub-millisecond echo times. At 3T, four echo-shifted coronal images were acquired with a minimum TE of 0.57 ms and an inter-echo spacing of 0.1 ms. Imaging was performed in free-breathing using three repetitions with a total scan time of 4.2 s. Subsequently, motion correction, field mapping, lung segmentation, and lung shim currents were calculated inline. Finally, the derived shim currents were saved locally, accessible by any custom sequence for application during scan preparation. The shimming method's performance was compared against the vendor's default settings in five healthy volunteers for functional lung imaging with matrix pencil decomposition using balanced steady-state free precession (bSSFP).
resultsField mapping and shimming of the lung was successfully performed in all volunteers and derivation of shim currents took less than 10s. At 3T, subject-specific shimming reduced the mean frequency offset in the lung by up to 45 Hz and the frequency range (max-min) by up to 180 Hz. Overall, this improved bSSFP signal homogeneity, resulting in more uniform functional images.
conclusionsThis work demonstrates a robust, automated shimming solution for lung imaging at 3T, easily integrable into clinical workflows. The technique significantly enhances image quality and reliability for high-field, bSSFP-based pulmonary imaging. At lower field strengths, the method can possibly relax TR constraints, reducing SAR and peripheral nerve stimulation.