Evidence map›Paper›PMID 41469873›Full record

ArticleBMC oral health2025

3D finite element analysis of removable partial denture with lingual bar major connector constructed with cobalt chromium, PEEK, PEKK, and BioHPP.

Aisha Zakaria Hashem Mostafa, Ahmed Sameh, Nermeen El Sayed El-Khamisy

Abstract readComparative Study
In one paragraph

Article in BMC oral health, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Aisha Zakaria Hashem MostafaProsthodontic Department, Faculty of Dentistry, Mansoura University, Mansoura, Egypt.
Ahmed SamehProduction Engineering and Mechanical Design Department, Faculty of Engineering, Mansoura University, Mansoura, Egypt. asameh192@mans.edu.eg.
Nermeen El Sayed El-KhamisyProsthodontic Department, Faculty of Dentistry, Mansoura University, Mansoura, Egypt.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectiveThe current study used finite element analysis (FEA) to investigate the rigidity, through Von-Mises stresses and deformation of cobalt-chromium, Polyetheretherketone (PEEK), Polyetherketoneketone (PEKK), and BioHPP constructed for mandibular class I RPD with lingual bar major connector. Although the general stress-strain relationships followed expected mechanical behavior, this study provides quantitative insight into how different polymeric materials (PEEK, PEKK, BioHPP) compare with conventional cobalt-chromium frameworks under identical loading conditions. MATERIALS AND

methodsFour virtual models for mandibular Kennedy's class I situations, with the remaining eight anterior and premolar teeth, were generated that closely resemble the actual cases. Mandibular cobalt-chromium class I RPD with lingual bar major connectors, half pear-shaped, with standard dimensions (50 × 5 × 3 mm), was constructed as a control group. Another 3 models similar to the control group in dimension but with 3 materials, PEEK, PEKK, and Bio HPP, were constructed as study groups. 3D finite element analysis was done, and data were collected and analyzed using bar chart.

resultsThe PEEK, PEKK, and Bio HPP major connectors showed more deformation (0.1798, 0.17554, 0.17104)mm, indicating weaker rigidity than the cobalt-chromium (0.028999)mm. The maximum von Mises stresses of the cobalt chromium lingual bar were 646.14 MPa, demonstrating stronger rigidity than the PEEK, PEKK, and Bio HPP, which exhibited values of 303.28, 295.29, and 301.23 MPa, respectively. In the case of PEEK and Bio HPP, greater stresses were transmitted to the supporting structures, particularly the mucosa, with maximum von Mises stress values of 3.0319 and 2.3047 MPa. These values were higher than the maximum stress transmitted to the mucosa in cobalt chromium, which was 0.60382 MPa. However, PEKK transmitted lower stresses to the supporting structures, particularly the mucosa, with a maximum von Mises stress value of 0.082096 MPa.

conclusionUsing the same size of PEEK, Bio HPP as the optimal dimensions of the cobalt chromium major connector of the mandibular class I RPD causes increased stress and deflection to the supporting structure, teeth, and soft tissues. This can be assigned to the PEEK's, Bio HPP, lower stiffness compared to cobalt chromium. On the other hand, PEKK showed decreased stresses and deflection of the supporting structure. When using PEEK and Bio HPP for class I RPD, the mandibular major connector must be larger.

Indexed as

Chromium AlloysDental MaterialsDenture DesignDenture, Partial, RemovableFinite Element AnalysisImaging, Three-DimensionalKetonesPolyethylene GlycolsBenzhydryl CompoundsBenzophenonesDental Stress AnalysisHumansPhenylenediaminesPolymersStress, MechanicalBenzhydryl CompoundsBenzophenonesChromium AlloysDental MaterialsKetonesPhenylenediaminespolyetheretherketonePolyethylene GlycolsPolymersBio HPPCobalt-chromiumDeflectionLingual barMajor connectorPolyetheretherketone (PEEK)Polyetherketoneketone (PEKK)RigidityRPDVon-Mises stress

Identifiers

PMID41469873
PMCPMC12857090

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390