Evidence map›Paper›PMID 41405045›Full record

ArticleThe journal of extra-corporeal technology2025

Endo-aortic balloon occlusion for reoperative minimally invasive mitral valve replacement on intra-aortic balloon pump: a case report.

Jenna L Cornibe, Vivek Patel, Austin Erney

Abstract readCase Reports
In one paragraph

Article in The journal of extra-corporeal technology, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Jenna L CornibeJupiter Medical Center, 1210 S Old Dixie Hwy, Jupiter, FL 33458, USA - Perfusion Services, Epic Specialty Staffing, 2250 McGregor Blvd Suite 3300, Fort Myers, FL 33901, USA.
Vivek PatelJupiter Medical Center, 1210 S Old Dixie Hwy, Jupiter, FL 33458, USA.
Austin ErneyJupiter Medical Center, 1210 S Old Dixie Hwy, Jupiter, FL 33458, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Minimally invasive approaches are the future of cardiac surgery. Coupled with that are devices such as endo-aortic balloon occlusion (EABO) which allow the success of such procedures. EABO has also proven invaluable in high-risk or reoperative cases. We report a severely high-risk patient who presented for reoperative minimally invasive mitral valve replacement with an intra-aortic balloon pump (IABP) and multiple high-dose vasopressors. The EABO was utilized over a transthoracic cross clamp (TCC). With careful maneuvering upon ascent into the aorta, the deflated endoballoon easily passed beyond the deflated IABP balloon under transesophageal echocardiogram (TEE) guidance. After completion of arrest, surgical placement of the bioprosthetic valve, and spontaneous return of cardiac rhythm once EABO was completed, the deflated endoballoon was carefully retracted back down through the aorta past the deflated IABP balloon. The integrity of the IABP balloon remained intact and was able to resume function once cardiopulmonary bypass (CPB) was terminated. We found that EABO was a safe and effective alternative technique for aortic occlusion in our patient supported on IABP with previous sternotomy.

Indexed as

Balloon OcclusionHeart Valve Prosthesis ImplantationIntra-Aortic Balloon PumpingMinimally Invasive Surgical ProceduresMitral ValveHumansMaleReoperationTreatment OutcomeEABOEndo-aortic balloon occlusionEndoballoonIABPMinimally invasive