Evidence map›Paper›PMID 41405034›Full record

ArticleThe journal of extra-corporeal technology2025

Determination of insensible water loss and sodium accumulation behavior from the Medtronic Nautilus Extracorporeal Membrane Oxygenation (ECMO) oxygenator: An in vitro study.

Carrie Whittaker Striker, Gabriel Kong

Abstract read
In one paragraph

Article in The journal of extra-corporeal technology, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors.

Carrie Whittaker StrikerDepartment of Cardiovascular Perfusion, Children's Wisconsin, Milwaukee, WI, USA - Electrical, Computer and Biomedical Engineering Department, Milwaukee School of Engineering, Milwaukee, WI, USA.ORCID 0009-0002-2744-7323
Gabriel KongElectrical, Computer and Biomedical Engineering Department, Milwaukee School of Engineering, Milwaukee, WI, USA.ORCID 0009-0006-9111-7416

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

introductionFluid and electrolyte balance is closely managed in extracorporeal membrane oxygenation (ECMO) patients. Neglecting oxygenator-related insensible fluid losses can distort fluid balance and electrolyte levels. While ECMO oxygenator insensible losses are reported, they remain undefined for the Medtronic Nautilus oxygenator. Through in vitro analysis of the Nautilus, we quantified insensible water losses while concurrently observing sodium behavior.

methodsInsensible water losses were determined using a closed circuit. A 12-hour pilot run was conducted to saturate the oxygenator and determine probable water loss rates and sodium accumulation behaviors. Fluid loss and sodium measurements were made at 0, 6, and 12 h. Immediately following the pilot run, three randomly assigned sweep gas rates, 0.5, 1.0, and 1.5 L/min, were evaluated over a 24 h period and replicated in triplicate. The circuit parameters were consistent and controlled for each trial. Data were collected at 0, 12, and 24 h for visualized water loss in the reservoir. After each trial, sterile water was introduced into the circuit via syringe and recorded as replacement volume. Sodium measurements were made for three trials (0.5, 1.0, and 1.5) and collected at 0 and 24 h.

resultsUsing linear regression analysis, the following insensible water loss equations were produced. Visualized volume: 2.74 mL/h per 1 L/min sweep gas rate or 65.66 mL/day per 1 L/min of sweep gas rate (p < 0.001). Replacement volume: 3.02 mL/h per 1 L/min of sweep gas rate or 72.5 mL/day per 1 L/min of sweep rate (p < 0.001). Sodium accumulation was observed, but not statistically significant due to the small sample size.

conclusionInsensible water loss in the Nautilus ECMO oxygenator increases linearly with sweep gas rate (p < 0.001), leading to sodium accumulation through evaporation. These losses and the associated risk for hypernatremia should be considered when managing a patient's fluid and electrolyte balance on ECMO.

Indexed as

Extracorporeal Membrane OxygenationOxygenators, MembraneSodiumWaterHumansWater-Electrolyte BalanceSodiumWaterECMOExtracorporeal membrane oxygenationFluid balanceHypernatremiaInsensible water lossOxygenatorSodium

Identifiers

PMID41405034
PMCPMC12710248

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.