Quantitative validation of data-driven motion correction for brain PET using phantom with motion generator system.

Yuto Kamitaka, Muneyuki Sakata, Keiichi Oda, Akie Katsuki, Hirofumi Kawakami, Kei Wagatsuma, Masato Kobayashi, Kenji Ishii

Abstract read
In one paragraph

Article in EJNMMI physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors.

Yuto KamitakaResearch Team for Neuroimaging, Tokyo Metropolitan Institute for Geriatrics and Gerontology, 35-2, Sakae-cho, Itabashi-ku, Tokyo, 173-0015, Japan.
Muneyuki SakataResearch Team for Neuroimaging, Tokyo Metropolitan Institute for Geriatrics and Gerontology, 35-2, Sakae-cho, Itabashi-ku, Tokyo, 173-0015, Japan. sakata@pet.tmig.or.jp.ORCID http://orcid.org/0000-0001-8161-0916
Keiichi OdaResearch Team for Neuroimaging, Tokyo Metropolitan Institute for Geriatrics and Gerontology, 35-2, Sakae-cho, Itabashi-ku, Tokyo, 173-0015, Japan.
Akie KatsukiGE HealthCare Japan, 4-7-127, Hino-shi, Tokyo, 191-8503, Japan.
Hirofumi KawakamiGE HealthCare Japan, 4-7-127, Hino-shi, Tokyo, 191-8503, Japan.
Kei WagatsumaResearch Team for Neuroimaging, Tokyo Metropolitan Institute for Geriatrics and Gerontology, 35-2, Sakae-cho, Itabashi-ku, Tokyo, 173-0015, Japan.
Masato KobayashiDivision of Health Sciences, Graduate School of Medical Sciences, Kanazawa University, 5-11-80, Kodatsuno, Kanazawa, 920-0942, Japan.
Kenji IshiiResearch Team for Neuroimaging, Tokyo Metropolitan Institute for Geriatrics and Gerontology, 35-2, Sakae-cho, Itabashi-ku, Tokyo, 173-0015, Japan.

Funding

GE Healthcare investigator-sponsored research grant
6 · The paper itself

Abstract

backgroundHead motion during brain positron emission tomography (PET) degrades image quality and quantitative accuracy. Therefore, a data-driven motion correction (MC) method utilizing ultrafast list-mode reconstruction technology has been proposed and shown to considerably improve image quality. However, reproducing accurate actual motions and motion-free images using clinical data alone remains challenging. This study aimed to quantitatively evaluate data-driven MC using a brain phantom for known tracer distributions and a custom-made motion generator system for variable known motions.

methodsHoffman 3D brain phantom was filled with 20 and 3 MBq of [

resultsThe motion generator system successfully reproduced the designed motions. Motion estimation remained stable under high-radioactivity condition but showed reduced stability under low-radioactivity condition, particularly with shorter frame durations. Under both conditions, longer frame durations led to underestimation of continuous motion. The data-driven MC improved %contrast and gray matter CV% across all conditions, with shorter frame durations providing better correction for quick or continuous motions. However, shorter frame durations increased statistical noise, especially under low-radioactivity condition.

conclusionThe data-driven MC effectively improved the quality of motion-affected PET images under both high- and low-radioactivity conditions, indicating its broad applicability. However, correction accuracy deteriorated under the lower-radioactivity condition.

Indexed as

[18F]FDGAlzheimer’s diseaseAmyloid PETData-driven motion correctionDementiaHead motionImage quality

Identifiers

PMID41400773
PMCPMC12824033

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.