Evidence map›Paper›PMID 41374655›Full record

ArticleSensors (Basel, Switzerland)2025

Measuring Relative Component Motion and Stability in Total Hip Replacements Using a Magnetic Position and Orientation Sensing System.

Oliver G Vickers, Peter R Culmer, Graham H Isaac, Robert W Kay, Matthew P Shuttleworth, Tim N Board, Sophie Williams

Abstract read
In one paragraph

Article in Sensors (Basel, Switzerland), 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Oliver G VickersThe Faculty of Engineering and Physical Sciences, University of Leeds, Leeds LS2 9JT, UK.ORCID 0000-0002-0682-5148
Peter R CulmerThe Faculty of Engineering and Physical Sciences, University of Leeds, Leeds LS2 9JT, UK.ORCID 0000-0003-2867-0420
Graham H IsaacThe Faculty of Engineering and Physical Sciences, University of Leeds, Leeds LS2 9JT, UK.ORCID 0000-0003-2613-3496
Robert W KayThe Faculty of Engineering and Physical Sciences, University of Leeds, Leeds LS2 9JT, UK.ORCID 0000-0002-2283-3789
Matthew P ShuttleworthThe Faculty of Engineering and Physical Sciences, University of Leeds, Leeds LS2 9JT, UK.ORCID 0000-0002-0538-6438
Tim N BoardThe Centre for Hip Surgery, Wrightington Hospital, Wigan WN6 9EP, UK.ORCID 0000-0002-3295-2087
Sophie WilliamsThe Faculty of Engineering and Physical Sciences, University of Leeds, Leeds LS2 9JT, UK.ORCID 0000-0002-6963-965X

Funding

Engineering and Physical Sciences Research Council EP/L014823/1Engineering and Physical Sciences Research Council EP/R003971/1
6 · The paper itself

Abstract

An instrumented total hip replacement (THR) implant capable of remote and continuous monitoring would be an attractive prospect for a surgeon to conveniently track the recovery of their patients. Measuring the relative motion of the prosthesis components would provide insight into joint kinematics and contribute to the detection of adverse events including impingement and subluxation. The aim of this study was to develop a sensing system to measure the relative orientation and translation of the prosthesis components. A tri-axis magnetometer and a permanent magnet were integrated into clinically available THR components, forming a magnetic position and orientation sensing system. A robotic arm was used to articulate the components through controlled motion routines and record the orientation of the components. The output of the robot arm and a camera tracking system were used to validate the performance of the sensing system. The sensing system measured the relative orientation of the components to two degrees of freedom with an RMSE of <4.0° and measured the displacement of the femoral head during an impingement-driven subluxation motion with an RMSE of 0.2 mm. This proof-of-concept work has shown that magnetic sensing technology can track the position and orientation of THR components. With further development, this sensing method could feature within an instrumented THR implant.

Indexed as

Arthroplasty, Replacement, HipHip ProsthesisBiomechanical PhenomenaHip JointHumansMotionRange of Motion, ArticularRoboticsinstrumented orthopaedic prosthesisjoint impingementmagnetic position and orientation sensing systemsubluxationtotal hip replacement

Identifiers

PMID41374655
PMCPMC12693991

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked