ArticleJournal of applied clinical medical physics2025
High-accuracy quality control method of CT system couch tops for treatment planning via an advanced 3D coordinate measuring machine.
Article in Journal of applied clinical medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- High-accuracy quality control method of CT system couch tops for treatment planning via an advanced 3D coordinate measuring machine.Journal of applied clinical medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
Abstract
backgroundQuality control (QC) methods for computed tomography (CT) systems used in treatment planning have not been updated since the release of the Task Group (TG) 66 report by the American Association of Physicists in Medicine (AAPM) in 2003. Conventional QC methods for CT systems fail to fulfill the requirements of high-precision radiation therapy. Moreover, because the geometric accuracy of CT systems can affect the accuracy of radiation therapy, which is particularly critical in high-precision radiation therapy, a highly accurate QC method is required. PURPOSE: This study aimed to develop a high-accuracy QC method for CT system couch tops suitable for high-precision radiation therapy by utilizing a wide-area three-dimensional (3D) coordinate measuring machine (3D-CMM), a type of laser tracker.
methodsWe used a 3D-CMM, which includes a wireless probe and a camera unit, focusing on a SOMATOM go.Open Pro CT system. The system was set up in accordance with the reference method outlined in the AAPM TG66 report guidelines. The initial phase verified the accuracy of 3D-CMM measurement within a CT room using a micrometer. Subsequently, a novel continuous measurement method was developed to enable the real-time tracking of the couch-top displacement during travel. This new method was evaluated against the standard manual measurement method. Measurements were performed at 0.1 s intervals at various index positions on the couch top, with and without added weight to simulate the presence of a patient. The gathered data were analyzed to assess the couch-top displacement, horizontality and orthogonality relative to the imaging plane, providing a comprehensive evaluation of the stability and alignment of the couch top.
resultsThe difference between the micrometer and measured shifts peaked at 0.08 mm. The continuous measurements agreed with the standard measurements within one standard deviation of the three measurements. The maximum displacements of the couch top were 5.23 and 2.00 mm in the vertical axis, with and without a weight load, respectively. There were differences in the displacement at each measurement point. In the lateral axis, the maximum displacements were 2.05 and 2.09 mm with and without a weight load, respectively. The maximum displacement in the imaging plane was observed at approximately half the distance traveled by the couch top. As the couch top traveled, the horizontal angle in the imaging plane of the couch top varied from 0.06° to 0.07° and 0.25° without a weight load and from 0.09° to 0.21° and 0.34° with a weight. The orthogonal angles of the couch top varied from 0.07° to 0.08° and 0.14° without a weight load and from 0.05° to 0.23° and 0.46° with a weight.
conclusionsThe developed QC method for the couch tops of CT systems can evaluate the displacement of the couch top and its horizontality and orthogonality to the imaging plane with detailed submillimeter and subdegree accuracy levels. The high-accuracy QC of CT systems can improve the accuracy of irradiation in radiation therapy.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials