Phantom-based quantification of the spectral accuracy of lower extremity vascular imaging using a second-generation dual-layer spectral CT at 100 kVp and 120kVp.
Article in BMC medical imaging, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors.
Funding
Abstract
backgroundThis study quantitatively evaluated the effects of tube voltage, CTDIvol, energy level, and iodine concentration on the performance of lower-extremity vascular imaging using a second-generation dual-layer spectral CT (DSCT). We further assessed the potential for reducing radiation dose and contrast agent use without compromising image quality.
methodsA spectral CT phantom with tissue-equivalent inserts and varying iodine concentrations (4, 8, 12, and 16 mgI/mL) was scanned at different tube voltages (100/120 kVp), CTDIvol levels (2.5, 5, 10 mGy), and virtual monoenergetic levels (40, 70, 100 keV), alongside conventional imaging. Vessel diameters of 2.5, 4, and 5 mm were evaluated. Objective image quality was assessed via CT attenuation, SD, SNR, and CNR. Two radiologists used a 5-point Likert scale to assess the subjective image quality.
resultsNo significant differences were observed in subjective or objective scores between tube voltages or across CTDIvol levels (p > 0.05). Low-energy 40 keV images yielded comparable quality to 70 keV and conventional images. At 40 keV, ultra-low iodine concentrations (4 mgI/mL) achieved image quality similar to higher concentrations (12/16 mgI/mL) in conventional scans. The combination of 100 kVp, 2.5 mGy, and 40 keV provided superior image quality relative to conventional CT.
conclusionsSecond-generation DSCT at 100 kVp provides high-accuracy lower-limb vascular imaging, even at low radiation doses and contrast concentrations, supporting its potential for clinical dose-reduction strategies.