Evidence map›Paper›PMID 41286510›Full record

ArticleMagnetic resonance in medicine2026

Read My Leads: Subject-Specific RF Hazard Assessment and Mitigation for DBS Implants in MRI.

Berk Silemek, Frank Seifert, Mevlüt Yalaz, Michael Höft, Günther Deuschl, Arzu Has Silemek, Rüdiger Brühl, Bernd Ittermann, Lukas Winter

Abstract read
In one paragraph

Article in Magnetic resonance in medicine, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors.

Berk SilemekPhysikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany.ORCID https://orcid.org/0000-0001-8227-3632
Frank SeifertPhysikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany.ORCID https://orcid.org/0000-0002-7065-2528
Mevlüt YalazDepartment of Electrical and Information Engineering, Christian-Albrechts-Universität zu Kiel, Kiel, Germany.ORCID https://orcid.org/0000-0001-6388-6847
Michael HöftDepartment of Electrical and Information Engineering, Christian-Albrechts-Universität zu Kiel, Kiel, Germany.ORCID https://orcid.org/0000-0001-9352-2868
Günther DeuschlDepartment of Neurology, Christian-Albrechts-Universität zu Kiel, Kiel, Germany.ORCID https://orcid.org/0000-0002-4176-9196
Arzu Has SilemekDepartment of Neurology, Cedars-Sinai Medical Center, Los Angeles, California, USA.ORCID https://orcid.org/0000-0003-4609-0956
Rüdiger BrühlPhysikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany.ORCID https://orcid.org/0000-0003-0111-5996
Bernd IttermannPhysikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany.ORCID https://orcid.org/0000-0002-4087-471X
Lukas WinterPhysikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany.ORCID https://orcid.org/0000-0002-4381-275X

Funding

Deutsche Forschungsgemeinschaft CRC 1261European Association of National Metrology Institutes (European Partnership on Metrology, co-financed by the European Union's Horizon Europe Research and Innovation Programme and by the Participating States) 21NRM05 STASIS
6 · The paper itself

Abstract

purposeTo develop and validate a framework for personalized, implant-specific MRI safety assessments using feedback from commercial deep brain stimulation (DBS) systems. To further use this framework to suppress RF-induced heating with minimum compromise in imaging performance.

methodsTwo off-the-shelf DBS implantable pulse generators and a commercial 8-electrode DBS lead were utilized for quantitative safety assessments. In controlled phantom experiments, (i) RF-induced voltages on the DBS lead, and (ii) temperature-dependent admittance/impedance changes in the tissue surrounding the lead's electrodes were quantified. This information was used to suppress implant-related RF heating by calculating implant-friendly imaging modes. Experimental conditions included excitations with different RF transmit coils (8-channel 3 T and 7 T head coils, 2-channel 3 T body coil), over 1000 different exposure scenarios, different implant configurations, and the use of external reference probes (

results

conclusionAll the information needed for full in situ control of implant heating in MRI can be read directly from the DBS device. This approach would improve both patient safety and image quality while simultaneously reducing workload and responsibilities of the clinical personnel.

Indexed as

Deep Brain StimulationElectrodes, ImplantedMagnetic Resonance ImagingBrainEquipment DesignHot TemperatureHumansPhantoms, ImagingRadio Wavesdeep brain stimulationimpedanceimplantsMR safetyRF heatingsensors

Identifiers

PMID41286510
PMCPMC12850618

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.