Evaluation of the Diagnostic Accuracy of the Quantitative Point-of-Care SD Biosensor Standard G6PD Test for Assessment of G6PD Deficiency in Infectious Diseases.
Article in International journal of laboratory hematology, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundG6PD deficiency affects about 500 million people worldwide and is prevalent in many malaria-endemic settings. People with G6PD deficiency are at risk of hemolysis when exposed to certain medications, including 8-aminoquinoline drugs used to treat Plasmodium vivax malaria. Increasing access to testing for G6PD deficiency at or near the point of care is critical to expanding the safe treatment of P. vivax malaria. We aimed to evaluate the performance of a semiquantitative test for G6PD deficiency, the SD Biosensor Standard G6PD Test with the Brewer's method, in a reference site for the treatment of infectious diseases.
methodWe evaluate the diagnostic accuracy performance of the SD Biosensor Standard G6PD Test in 125 individuals with infectious diseases and other illnesses. MAIN
resultsWe observed a trend of concordance between the G6PD status (Deficient or Normal), with low frequencies of discordances in the screenings. The strength of agreement between G6PD tests was classified as almost perfect in all participants (k = 0.82, 95% CI = 0.66, 0.97) and in sex subgroups: females (k = 0.83, 95% CI = 0.59, 1.00) and males (k = 0.81, 95% CI = 0.59, 1.00). The total concordance percentage (number of concordances/total) was 95% for females, 97% for males, and 96% overall.
conclusionThe SD Biosensor STANDARD G6PD test is an innovative point-of-care solution. It offers quantitative measurement of glucose-6-phosphate dehydrogenase activity while normalizing for hemoglobin levels. This advancement enables its use in lower-tier clinical and laboratory settings, expanding access to accurate G6PD testing.