An automated daily QA strategy for dosimetry and imaging guidance check of a novel PET/CT ring gantry linac.
Article in Journal of applied clinical medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- An automated daily QA strategy for dosimetry and imaging guidance check of a novel PET/CT ring gantry linac.Journal of applied clinical medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThe RefleXion X1 is a novel PET/CT-guided ring gantry linac that introduces unique geometry and workflow compared to conventional systems. Existing QA guidelines (AAPM TG-142, 148, 306) cover standard linacs and tomotherapy but do not address this platform's needs. Therefore, an automated daily QA solution is required to ensure accurate output, imaging guidance, and couch positioning, which motivated the development and evaluation of the workflow presented in this study. PURPOSE: To develop and evaluate an automated, efficient, and comprehensive daily quality assurance (QA) workflow for a novel PET/CT-guided ring gantry linear accelerator (RefleXion X1), incorporating tests for dosimetry, imaging guidance, and couch positioning accuracy.
methodsAn IMRT QA plan was designed following the CT scan of the TomoDose 2D diode detector array (Sun Nuclear, Melbourne, FL) within the RefleXion treatment planning system (TPS). Several spherical targets of varying sizes were contoured to conform the dose distribution to these targets and evaluate beam output, laser localization, profile constancy, and CT-MV isocenter coincidence. Linear models correlating spatial location of dose peak with intentional shifts in different directions were developed and integrated into an in-house analysis software, iQA, written in C#. The software was validated for robustness and successfully implemented in clinical practice.
resultsValidation with 42 intentional shifts showed submillimeter accuracy, with mean deviations of 0.2 mm (X), 0.1 mm (Y), and 0.79 mm (Z), and R
conclusionThis fully automated, single-delivery QA workflow enables efficient and robust evaluation of multiple IGRT performance parameters. It improves standardization, reduces human error, and reliably monitors long-term machine stability. The tools and methodology may be adopted by other clinics seeking a practical and automated daily QA solution. Future developments will focus on incorporating PET QA and rotational couch validations to further enhance the system's comprehensiveness and adaptability.