ArticleJournal of applied clinical medical physics2025
Commissioning of a combined stereoscopic x-ray and surface guided radiotherapy system for performing breath hold gated lung radiotherapy.
Article in Journal of applied clinical medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- Commissioning of a combined stereoscopic x-ray and surface guided radiotherapy system for performing breath hold gated lung radiotherapy.Journal of applied clinical medical physics · 2025Article
- Surface-guided Radiation Therapy in Clinical Practice: A Review with Perspectives on Implementation in Resource-constrained Settings.Journal of medical physicsReview
- Optimizing Computed Tomography-derived Surface Monitoring for Accurate Cancer Radiotherapy Setup.In vivo (Athens, Greece)Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
2 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundExacTrac Dynamic (ETD) is a novel patient positioning and motion management system for radiation therapy that combines surface guidance, thermal imaging, and stereoscopic x-rays for image guidance. This combination of technology enables both intrafraction monitoring and surface-based breath hold gating capabilities. PURPOSE: The purpose of this study was to use the American Association of Physicists in Medicine (AAPM) TG-302 framework to commission two independent ETD systems for performing breath hold gated lung radiotherapy.
methodsA series of static and dynamic tests were performed using an anthropomorphic thorax phantom placed on a dynamic motion platform. Static tests included localization accuracy and reproducibility, and the impact of reference surface selection on localization. Spatial drift was evaluated during system warmup, during typical treatment session duration, and with changing temperature. Dynamic tests assessed localization accuracy and reproducibility, as well as gating latency. Finally, end-to-end testing of a simulated breath hold lung stereotactic body radiation therapy (SBRT) treatment was performed, with both gated and ungated dose measurements being acquired.
resultsStatic and dynamic localization accuracy averaged less than 1 mm, with reproducibility within 0.4 mm. An optimal Hounsfield Unit (HU) threshold of-700 for the reference surface contour was demonstrated to have the best localization accuracy. The maximum observed spatial drift for all tests was less than 1 mm. Congruence between surface guidance, x-ray imaging, and CBCT were all within 1 mm. Mean beam on and off latency averaged 1023 and 527.1 ms, respectively. End-to-end testing demonstrated less than 1% dose difference between gated and non-gated treatment delivery. The independent ETD systems demonstrated consistent performance across all commissioning tests.
conclusionsAll commissioning tests were within acceptable tolerances published in the AAPM TG-302 report. This study demonstrated that the ETD system is suitable for performing surface-guided breath hold gated lung SBRT treatments.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.