Performance evaluation of the TE Air wireless handheld ultrasound in cardiac applications: a prospective comparative study.

Beibei Ge, Mingxia Li, Hanlin Cheng, Zhanru Qi, Xiaoxian Wang, Fen Chen, Zhongqing Shi, Guanjun Guo, Chunjie Shan, Jinyang Qi and 3 more

Abstract read
In one paragraph

Article in The ultrasound journal, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

13 authors.

Beibei Ge *Department of Ultrasound Medicine, Nanjing Drum Tower Hospital, Affiliated Hospital of Medical School, Nanjing University, Nanjing, 210008, People's Republic of China.
Mingxia Li *Department of Ultrasound Medicine, Nanjing Drum Tower Hospital, Affiliated Hospital of Medical School, Nanjing University, Nanjing, 210008, People's Republic of China.
Hanlin ChengSchool of Biological Sciences and Medical Engineering, Southeast University, Nanjing, 210096, People's Republic of China.
Zhanru QiDepartment of Ultrasound Medicine, Nanjing Drum Tower Hospital, Affiliated Hospital of Medical School, Nanjing University, Nanjing, 210008, People's Republic of China.
Xiaoxian WangDepartment of Ultrasound Medicine, Nanjing Drum Tower Hospital, Affiliated Hospital of Medical School, Nanjing University, Nanjing, 210008, People's Republic of China.
Fen ChenDepartment of Ultrasound Medicine, Nanjing Drum Tower Hospital, Affiliated Hospital of Medical School, Nanjing University, Nanjing, 210008, People's Republic of China.
Zhongqing ShiDepartment of Ultrasound Medicine, Nanjing Drum Tower Hospital, Affiliated Hospital of Medical School, Nanjing University, Nanjing, 210008, People's Republic of China.
Guanjun GuoDepartment of Ultrasound Medicine, Nanjing Drum Tower Hospital, Affiliated Hospital of Medical School, Nanjing University, Nanjing, 210008, People's Republic of China.
Chunjie ShanSchool of Biological Sciences and Medical Engineering, Southeast University, Nanjing, 210096, People's Republic of China.
Jinyang QiDepartment of Ultrasound Medicine, Yancheng First People's Hospital Affiliated of Nanjing University Medical College, Nanjing University, Yancheng, 224000, People's Republic of China.
Shouhua LuoSchool of Biological Sciences and Medical Engineering, Southeast University, Nanjing, 210096, People's Republic of China.
Fang Xu *Department of Ultrasound Medicine, Nanjing Drum Tower Hospital, Affiliated Hospital of Medical School, Nanjing University, Nanjing, 210008, People's Republic of China. snowmeimei@yeah.net.
Jing Yao *Department of Ultrasound Medicine, Nanjing Drum Tower Hospital, Affiliated Hospital of Medical School, Nanjing University, Nanjing, 210008, People's Republic of China. jingyao@nju.edu.cn.

Funding

National Natural Science Foundation of China No.81771844National Natural Science Foundation of China No.82371981
6 · The paper itself

Abstract

aimTo evaluate the reliability and reproducibility of the TE Air wireless handheld ultrasound device in clinical cardiac imaging by comparing its performance with a high-end reference system.

methods161 patients for good-quality echocardiographic images were included in this prospective study. Each patient underwent sequential imaging using both the TE Air device (Mindray) and the high-end reference device (Philips EPIQ 7 C). Nine standard cardiac views were acquired. Image quality was assessed manually by two blinded echocardiographers and via proprietary AI software, respectively. The following key parameters were analyzed basing on the images: diastolic thickness of interventricular septal (IVSTd) and left ventricular posterior wall (LVPWTd), left ventricular end-diastolic (LVDd) and end-systolic diameter (LVDs), aortic diameter (AOD), left atrial anteroposterior diameter (LAD), Early (E) and late (A) diastolic velocities of the mitral valve in PW mode, as well as early diastolic velocities at the septal (EmS) and lateral (EmL) mitral annulus. Regional wall motion abnormality (RWMA), bicuspid aortic valve (BAV), atrial septal defect (ASD), left ventricular ejection fraction (LVEF) and valvular regurgitation degree were independently evaluated.

resultsThe TE Air demonstrated comparable image quality to the high-end reference system in both manual (64.95 ± 1.24 vs. 64.19 ± 1.63, P = 0.28) and AI-based evaluations (65.07 ± 1.02 vs. 63.80 ± 1.68, P = 0.06). Structural measurements showed high inter-device consistency, with ICCs of 0.77/0.74 for IVSTd/LVPWTd, 0.95/0.96 for LVDd/LVDs, and 0.82/0.98 for AOD/LAD (all P < 0.001). Functional parameters also demonstrated strong agreement (ICC: 0.91/0.92 for mitral E/A waves; 0.79/0.85 for EmS/EmL; P < 0.001). The TE Air had sensitivities of 81.8% for RWMA, 100% for ASD and BAV, and 93.5% for LVEF < 50%. Diagnostic agreement was excellent for LVEF (κ = 0.96, P < 0.001) and valvular regurgitation (weighted κ = 0.89, P < 0.001).

conclusionThe TE Air wireless handheld ultrasound device exhibits high agreement with high-end reference device in image quality, measurements, and clinical diagnoses, supporting its potential for widespread use in point-of-care ultrasound (POCUS) clinical applications.