Optimizing dosimetric leaf gap parameters for radiation therapy: A comparative study using three diode-array verification tools and an ionization chamber.
Article in Journal of applied clinical medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Optimizing dosimetric leaf gap parameters for radiation therapy: A comparative study using three diode-array verification tools and an ionization chamber.Journal of applied clinical medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThe dosimetric leaf gap (DLG) value is a key parameter that affects dose calculation accuracy in radiotherapy when using the Eclipse treatment planning system. Diode array dosimeters are used to determine the optimal DLG value because they are convenient and fast, providing instant readout of planar dose distributions rather than a point dose. However, variations in gamma pass rates (GPRs) across devices, even for the same radiotherapy treatment plan, may lead to different DLG determinations. PURPOSE: To compare the optimal DLG values determined using three -diode array dosimeters with those obtained using an ionization chamber (IC) and to evaluate whether diode array dosimeters can reliably reproduce IC-based DLG values.
methodsOptimal DLG values were determined for 6 MV, 10 MV, and flattening filter-free (FFF) beams using ArcCHECK, Delta4, and SRS MapCHECK. The results were compared with the IC-based values across all energies. Radiochromic films validated the IC-based DLG values. Twenty volumetric modulated arc therapy plans were recalculated using ten different DLG values. The results were compared with the measured data. Optimal DLG values were determined using global gamma criteria (3%/2 mm and 2%/2 mm), local gamma (1%/1 mm), and dose difference criteria.
resultsVerification with radiochromic film using IC-based optimal DLG values yielded an average GPR of 98.7 ± 2.3% under the 3%/3 mm criterion. The average DLG difference between the IC and diode array dosimeters was -0.35 ± 0.23 mm (range: -0.80 to 0.26 mm), based on various gamma and dose difference criteria. A DLG difference of -0.35 mm showed dose deviations ranging from 0.27% to 1.05% across all beams. Except for one criterion, no significant differences were observed.
conclusionsDLG differences between IC and diode array dosimeters resulted in approximately 1% dose discrepancies, which are within the clinically acceptable range. This finding indicates that ArcCHECK, Delta4, and SRS MapCHECK can determine DLG values comparable to those obtained with IC.