Evidence map›Paper›PMID 41141229›Full record

ArticlePhysics and imaging in radiation oncology2025

A phantom study of internal target volume and mid-position accuracy in adaptive and conventional four-dimensional computed tomography across regular and irregular motion.

Bart J J Kremers, Dave S C van Gruijthuijsen, Dominique Reijtenbagh, Jacco L G Steenhuijsen, Mariska de Smet, Rob H N Tijssen

Abstract read
In one paragraph

Article in Physics and imaging in radiation oncology, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Bart J J KremersDepartment of Radiation Oncology, Catharina Hospital Eindhoven, Eindhoven, Netherlands (the).
Dave S C van GruijthuijsenDepartment of Radiation Oncology, Catharina Hospital Eindhoven, Eindhoven, Netherlands (the).
Dominique ReijtenbaghDepartment of Radiation Oncology, Catharina Hospital Eindhoven, Eindhoven, Netherlands (the).
Jacco L G SteenhuijsenDepartment of Radiation Oncology, Catharina Hospital Eindhoven, Eindhoven, Netherlands (the).
Mariska de SmetDepartment of Radiation Oncology, Catharina Hospital Eindhoven, Eindhoven, Netherlands (the).
Rob H N TijssenDepartment of Radiation Oncology, Catharina Hospital Eindhoven, Eindhoven, Netherlands (the).

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

This technical note evaluates the performance of an adaptive four-dimensional computed tomography (4DCT) acquisition method compared to conventional 4DCT using a motion phantom. Metrics assessed include deviations in volume, CT number, diameter, peak-to-peak amplitude and determination of the internal target volume (ITV) and mid-position. Under regular breathing, most measurements fall within predefined clinical tolerances for all systems. Under irregular motion, the adaptive method showed reduced deviations in ITV and minimal impact on mid-position determination. These findings support the clinical value of adaptive 4DCT in improving motion management and target definition accuracy in radiotherapy planning.

Indexed as

4DCT quality assuranceAdaptive 4DCTMid-position

Identifiers

PMID41141229
PMCPMC12547255

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.