ArticleEuropean journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery2025
Low Driveline Infection Rates in Patients With a Novel Fully Magnetically Levitated Ventricular Assist Device.
Article in European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT06878456 (Incidence and Management of Driveline Infection in Patients with CH-VAD), which is not on this map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Incidence and Management of Driveline Infection in Patients with CH-VAD: a Multi-centre, Observational Study
Who cites it
2 citing papers in PubMed.
- Technological advances and clinical application of left ventricular assist devices in the treatment of advanced heart failure.Journal of cardiothoracic surgery · 2026Review
- Worldwide research trends on left ventricular assist devices for heart failure: a bibliometric analysis.Journal of cardiothoracic surgery · 2026Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
16 authors.
Funding
Abstract
objectivesDriveline infection remains a major complication of long-term ventricular assist device therapy. Previous studies have associated the risk of infection with driveline design and physical properties. The CH-VAD is a fully magnetically levitated pump with an ultrathin (diameter 3.3 mm), highly flexible driveline designed to reduce infection risk. This study was conducted to assess the incidence, risk factors, and approaches to prevention and management of driveline infection in CH-VAD patients (NCT06878456).
methodsAll patients who underwent CH-VAD implantation between June 2017 and October 2024 at 9 centres were retrospectively analysed. Data related to clinical characteristics, surgical procedures, and driveline infection prevention and management details were collected, and risk factors were analysed.
resultsA total of 181 consecutive patients were included, with a median age of 56 years and 86% male. During a median support duration of 1.6 years, driveline infections occurred in 20 out of 181 patients (11.0%). Twenty-six events were observed, corresponding to a rate of 0.07 events per patient-year. Cumulative incidence of the first driveline infection was 2.8% at 6 months, 5.6% at 12 months, and 8.5% at 24 months. Frequent dressing changes were found to be significantly associated with increased driveline infection risk (HR 3.773, P = .022).
conclusionsPatients treated with CH-VADs experienced an encouraging low rate of driveline infection, considered among the lowest reported for durable ventricular assist devices. These findings have implications for both clinical management and driveline design.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.