Extravascular implantable cardioverter and leadless pacemaker interactions.

Alfonso Aranda Hernandez, G Stuart Mendenhall

Abstract read
In one paragraph

Article in Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Review
  2. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors.

Alfonso Aranda HernandezMedtronic Research and Technology, 8200 Coral Sea Street NE, Mounds View, MN 55112, USA.ORCID 0000-0003-0062-2865
G Stuart MendenhallScripps Memorial Hospital-La Jolla, La Jolla, CA, USA.ORCID 0000-0003-3208-8679

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

aimsTraditional cardiac pacing and defibrillation devices rely on leads connected to a subcutaneous pulse generator, which can result in complications such as vascular damage, infection, or lead failure. Advances in technology have led to the development of leadless pacemakers, which combine the battery, circuitry, and electrodes into a single self-contained unit, and extravascular implantable cardioverter-defibrillators (ICDs), which position electrodes outside the vasculature. These innovations offer promising alternatives for patients requiring both defibrillation and pacing, particularly those unable or unwilling to accommodate traditional leads. However, the interactions between extravascular ICDs and leadless pacemakers remain largely unexplored and currently lack regulatory approval for combined use. This study evaluates the interactions between leadless pacemakers and extravascular ICDs to assess their simultaneous operation. METHODS AND

resultsIn-silico simulations, saline-tank experiments, and Monte Carlo simulations were conducted to evaluate device interactions, focusing on ventricular fibrillation (VF) detection during pacing conditions. Ventricular fibrillation detection was unaffected by pacing pulse widths ≤0.24 ms, with a pacing pulse-to-VF amplitude ratio of ≤2 considered safe. Wider pulse widths or higher outputs progressively increased the risk of VF undersensing. Experiments confirmed that pacing pulses ≤3 V and ≤0.24 ms minimally impacted VF detection. Proximity of device affected pacing pulse amplitude sensed by the ICD, but pacemaker orientation did not. Monte Carlo simulations indicated a 0-4% probability of undesired interactions under clinically relevant conditions.

conclusionExtravascular ICDs and leadless pacemakers may safely coexist, with a low observed risk of VF undersensing in our study. Further clinical studies are needed to confirm these findings.

Indexed as

Cardiac Pacing, ArtificialDefibrillators, ImplantableElectric CountershockPacemaker, ArtificialVentricular FibrillationComputer SimulationEquipment DesignHumansModels, CardiovascularMonte Carlo MethodExtravascular ICDLeadless pacemakerVF detection

Identifiers

PMID41102868
PMCPMC12581963

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.