Echocardiographic assessment of a novel blood-guiding accessory for left ventricular assist devices in swine.
Article in Interdisciplinary cardiovascular and thoracic surgery, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- In Vivo Testing of a Second-Generation Prototype Accessory for Single Transapical Left Ventricular Assist Device Implantation.Bioengineering (Basel, Switzerland) · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectivesLeft ventricular assist devices are part of the foundations of mechanical circulatory support. As such devices evolve in technology, so does their implantation techniques, becoming less invasive. We developed a novel blood-guiding accessory for less invasive implantation, redirecting the blood intracardially into the ascending aorta through a transaortic outflow.
methodsTen healthy female swine (German Landrace, 104.4 ± 13.0 kg) underwent epicardial echocardiography before and after left ventricular assist device implantation. Cardiac geometry, aortic valve function and outflow positioning, were assessed before and after implantation.
resultsEpicardial echocardiography revealed sufficient biventricular unloading after device implantation. The end-diastolic and end-systolic left ventricular diameters decreased by 16% and 20%, respectively. Retrograde unloading was displayed by smaller diameters in left and right atria at 22% and 25%, respectively. At baseline, one animal presented mild aortic regurgitation, unchanged under mechanical support, whereas one developed mild de novo insufficiency. The outflow graft was either well centered within the aortic valve (n = 3) or between two cusps.
conclusionsA left ventricular assist device with the accessory allowed sufficient ventricular unloading, preserving biventricular function. Although the outflow graft was well centered within the aortic valve in only 30% of the cases, no relevant aortic regurgitation was found in the acute setting. Chronic testing in larger samples is required to analyze results in long term.