Diagnostic performance of point-of-care bilirubin testing with Bilistick 2.0 device at a South Indian clinical site.
Article in Pediatric research, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundNeonatal hyperbilirubinemia is common in India, yet many public hospitals lack access to reliable and timely bilirubin testing. We evaluated the diagnostic performance of a point-of-care (POC) bilirubin device, Bilistick 2.0, in a high-volume neonatal setting.
methodsWe conducted a prospective diagnostic accuracy study at Madras Medical College, Chennai. Neonates ≥28 weeks and ≤10 days old with jaundice or at high risk were enrolled. Paired TSB samples (Bilistick and reference lab) were collected from 153 neonates. Bland-Altman analysis, correlation, and subgroup comparisons were performed.
resultsOf 153 paired samples, 137 non-hemolyzed pairs were analyzed. Mean TSB was 14.39 ± 4.03 mg/dL (Bilistick) vs. 14.27 ± 3.73 mg/dL (reference). Mean absolute difference was 1.35 mg/dL; 79% of values were within ±2 mg/dL. Mean bias was +0.13 mg/dL. Correlation was r = 0.88 overall, and r = 0.96 for TSB > 20 mg/dL. Test failure rate was 10.5%, mostly during the initial learning phase.
conclusionBilistick 2.0 showed strong diagnostic agreement with reference laboratory bilirubin testing. With training and basic laboratory coordination, it may support safe, rapid bilirubin management in resource-limited neonatal care. IMPACT: The Bilistick 2.0 point-of-care device showed strong diagnostic agreement with reference laboratory bilirubin testing. Performance remained reliable in neonates with clinically significant bilirubin levels, especially >20 mg/dL. Turnaround time was under 15 min, and user-related test failures decreased after brief training. The device has potential for decentralized bilirubin monitoring in public-sector neonatal care in India.