Worldwide data analysis of BIOTRONIK ICD and CRT-D battery longevity.
Article in Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Cardiac Resynchronization Therapy With Defibrillator Using the JROAD-DPC Database: Cost-Effectiveness Analysis.Interactive journal of medical research · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundBattery longevity is a key consideration in the selection of implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds), with implications for both clinical outcomes and healthcare costs. A recent study suggested shorter-than-expected battery life for BIOTRONIK devices, prompting a comprehensive internal analysis. This analysis aimed to evaluate the real-world battery longevity of BIOTRONIK high-voltage devices (ICDs and CRT-Ds) across three generations, using global remote monitoring data.
methodsA total of 215,471 BIOTRONIK devices implanted between 2006 and 2024 were analyzed using data from the BIOTRONIK Home Monitoring Service Center. Devices were grouped into three generations (BIO 2000, 2010, 2020) and stratified by type: single-chamber ICD (VR), dual-chamber ICD (DR), CRT-D, and single-lead ICD with an atrial floating dipole (DX). Kaplan-Meier survival analysis was used to estimate median service times, with extrapolation applied to censored data.
resultsMedian service times increased substantially across generations (p < 0.001): VR devices improved from 92 (interquartile range 72-106) to 175 (169-182) months (+89%), DR from 73 (63-93) to 155 (132-168) months (+108%), CRT-D from 72 (62-79) to 111 (95-126) months (+52%), and DX from 97 (92-101) to 170 (163-175) months (+74%). DX devices demonstrated superior longevity compared to DR devices.
conclusionBattery longevity of BIOTRONIK ICDs and CRT-Ds has improved markedly with each generation. Newer devices demonstrate service times at or above industry standards, supporting their clinical and economic value.