Evidence map›Paper›PMID 41049310›Full record

ArticleAsiaIntervention2025

Clinical evaluation of the Hydra self-expanding THV: 30-day results from the GENESIS-II study.

Nagendra Boopathy Senguttuvan, John Jose, Anmol Sonawane, Sandeep Bansal, Rahul Gupta, Praveen Chandra

Registry-linked trialAbstract read
In one paragraph

Article in AsiaIntervention, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT07566624 (Spanish Prospective Registry of Hydra THV for Transcatheter Aortic Valve Implantation), which is not on this map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

NCT07566624 recruitingnot on this mapstarted 2026, after this paper: background citation

Spanish Prospective Registry of Hydra THV for Transcatheter Aortic Valve Implantation

Typeobservational_patient_registrySponsorFundación EPICRan2026 to 2029Enrolled100ConditionsSevere Aortic Valve Stenosis, Valvular Heart Disease, Aortic Valve StenosisArmsHydra THV
3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Nagendra Boopathy SenguttuvanDepartment of Cardiology, Sri Ramachandra Institute of Higher Education and Research, Chennai, India.
John JoseDepartment of Cardiology, Christian Medical College Vellore, Vellore, India.
Anmol SonawaneDepartment of Cardiology, Breach Candy Hospital Trust, Mumbai, India.
Sandeep BansalDepartment of Cardiology, Safdarjung Hospital, New Delhi, India.
Rahul GuptaDepartment of Cardiology, Apollo Hospitals Enterprise Limited, Thane, India.
Praveen ChandraDepartment of Cardiology, Medanta - The Medicity Hospital, Gurgaon, India.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: The Hydra transcatheter heart valve (THV) utilised in the GENESIS-II study is equipped with an active release mechanism for tentacle deployment, distinguishing it from the Hydra THV used in earlier studies. Aims: The primary objective of the GENESIS-II study is to assess the ongoing safety and performance of the Hydra THV in treating severe aortic stenosis in patients deemed at high surgical risk. Methods: This is a prospective, multicentre, non-randomised, investigational study conducted between November 2021 and November 2023. The study enrolled a total of 40 patients exhibiting high surgical risk and symptomatic severe aortic stenosis from 19 sites across India. The primary safety endpoint of the study was cardiovascular mortality at 30 days, while the primary performance endpoint was device success as defined by Valve Academic Research Consortium-2. The 6-month follow-up of the GENESIS-II study is ongoing. Results: Among the 40 patients, the average age was 74.4±6.7 years, and 65% were male. The rate of device success was 95%. At 30 days, 2.5% of patients experienced cardiovascular mortality, and the incidence of new permanent pacemaker implantation was 7.9%. There was a progressive enhancement in the effective orifice area, transitioning from 0.6±0.2 cm² at baseline to 1.9±0.6 cm² at 30 days (p<0.001). Similarly, the mean aortic valve gradient demonstrated significant improvement from 53.1±12.5 mmHg at baseline to 8.4±4.0 mmHg at 30 days (p<0.001) post-intervention. Conclusions: The Hydra THV is the first device with an active release mechanism for deploying the supra-annular valve. The observed high rates of device success and procedural safety in the GENESIS-II study may support the continued safety and performance of the device.

Identifiers

PMID41049310
PMCPMC12494087

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.