ArticleJournal of contemporary brachytherapy2025

X-ray spectra emitted by the electronic brachytherapy unit Papillon 50 measured with a CdTe spectrometer.

Jaroslav Šolc, Gustavo Kertzscher, Jana Šmoldasová, Vladimír Sochor, Peter Georgi

Abstract read
In one paragraph

Article in Journal of contemporary brachytherapy, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Jaroslav ŠolcCzech Metrology Institute, Brno, Czech Republic.
Gustavo KertzscherDepartment of Clinical Medicine, Aarhus University, Aarhus, Denmark.
Jana ŠmoldasováCzech Metrology Institute, Brno, Czech Republic.
Vladimír SochorCzech Metrology Institute, Brno, Czech Republic.
Peter GeorgiDepartment of Clinical Medicine, Aarhus University, Aarhus, Denmark.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Purpose: The aim of this paper was to present the measured photon fluence spectra emitted by the electronic brachytherapy (eBT) unit Papillon 50. Material and methods: Spectra were collected with a compact CdTe spectrometer in air at a distance of 52 cm from the unit output window, for three different applicators attached to the unit and at several angles with respect to the beam axis. Results: The fluence spectra recorded for this device showed lower energy of the photon spectrum end-point than anticipated. The end-point energy was found to be (47.6 ±0.2) keV for the nominal high voltage bias of 50 kV set by the operator. Full-energy peaks in the spectra identified characteristic photons emitted by the rhenium anode, and by molybdenum and silver supposedly present in the inner coating of the applicators. An asymmetry in the half-value layer values of measured spectra was observed. Conclusions: The presented work provides a comprehensive dataset of photon fluence spectra for traceable dosimetry using the Papillon 50 eBT device, which is required, for example, for correcting the energy dependence of detectors or as an input for Monte Carlo calculations. Spectrometry and dose characterization studies are needed for eBT systems.

Indexed as

CdTe detectorPapillon 50photon fluence spectrumPRISM-eBTX-ray beam spectrometry