Three-dimensional proton FLASH dose rate measurement at high spatiotemporal resolution using a novel multi-layer strip ionization chamber (MLSIC) device.
Article in Medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Pulse-by-pulse ultra-high resolution scintillation imaging of proton FLASH beams produced by a gantry-mounted synchrocyclotron.Physics in medicine and biology · 2025Article
- Emerging strategies in radiation therapy: promises and challenges of spatial fractionation, ultra-high dose rates, and nanoparticles.Journal of physics D: Applied physics · 2025Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
14 authors.
Funding
Abstract
backgroundCurrently, proton therapy is the main radiation treatment modality that can treat deeply seated targets at ultra-high dose rates. The safe translation of FLASH RT into clinic requires dedicated dosimeters capable of measurements at sufficiently high spatiotemporal resolution. PURPOSE: The objective of this work is to demonstrate the feasibility of three-dimensional (3D) measurements of dose rate and dose for FLASH pencil beam scanning (PBS) proton therapy.
methodsA multi-layer strip ionization chamber (MLSIC) device, along with a reconstruction algorithm, was designed and developed to reconstruct dose and dose rate distribution over a 3D volume. Our MLSIC is composed of 66 layers of strip ionization chamber arrays with total water-equivalent thickness (WET) of 19.2 cm along the beam direction. The first two layers, composed of 128 channels with orthogonal direction with respect to each other, provide the (x,y) coordinate. The other 64 layers contain 32 channels with 8 mm lateral spacing. Data readout at a high-speed of 6250 fps allows spot-by-spot measurement. To prove the concept, PBS proton therapy plans were delivered at conventional and FLASH dose rates. Dose and dose rate information were reconstructed in 3D using an in-house reconstruction algorithm.
resultsIon recombination remained under 1% in the majority of cases. 3D dose reconstruction showed agreement with the treatment planning software; the 3D gamma analysis of the reconstructed dose showed 96.2% (5 mm/5%) and 86.8% (3 mm/3%) passing rates with 10% threshold for the conventional dose rate plan, 99.1% (5 mm/5%) and 92.6% (3 mm/3%) passing rates for the FLASH dose rate plan. 3D dose rate distributions were successfully generated using different definitions.
conclusionsOur MLSIC device allows obtaining 3D dose and dose rate distribution of PBS proton beams at FLASH dose rates with high spatiotemporal resolution.