ArticleJournal of robotic surgery2025
Effectiveness of haptic feedback during local and remote robotic surgery: single-blind cadaveric study.
Article in Journal of robotic surgery, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed, 1 synthesis or guideline pooled it.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundRobotic surgery has rapidly evolved, with telesurgery emerging as a promising extension. However, the lack of haptic feedback remains a key limitation, potentially compromising surgical safety. This study aimed to evaluate the impact of haptic feedback on surgical performance in both local and remote robotic settings using a cadaver model.
methodsSix gastrointestinal surgeons were assigned to local or remote groups and performed standardized bowel traction tasks using the Saroa™ surgical robot. Haptic feedback was tested at three levels: none (0), moderate (0.5), and full (1.0). Each participant completed the task three times under each condition. Task completion time, grip force, and forceps path length were measured.
resultsGrip force significantly decreased with increasing haptic feedback levels, particularly in the non-dominant (left) hand. No significant differences in task completion time or forceps path length were found across feedback levels or between local and remote settings, except for a longer right-hand path length in the remote group at feedback level 0.5.
conclusionsHaptic feedback reduced grip force in both local and remote robotic surgery, suggesting enhanced surgical safety. However, its effect on efficiency metrics was limited. These findings support the integration of haptic feedback in telesurgical systems, although further validation with complex procedures and more participants is warranted.